Bard StatLock Intravenous (IV) Catheter Stabilization Kit, Catalog number: IVK0124, C.R. Bard, Inc...

C.R. Bard Inc / initiated 2009-04-02 / Terminated / root cause: manufacturing process

C.R. Bard, Inc., Urological Division began a recall of Bard StatLock Intravenous (IV) Catheter Stabilization Kit, Catalog number: IVK0124, C.R. Bard, Inc., Covington, GA 30014, USA on . The reason FDA records is "The affected units may contain extension sets with the microbore tubing instead of extension sets with the macrobore tubing". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1539-2009
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
C.R. Bard Inc
Product
Bard StatLock Intravenous (IV) Catheter Stabilization Kit, Catalog number: IVK0124, C.R. Bard, Inc., Covington, GA 30014, USA.
Product code
FPA Set, Administration, Intravascular
Reason for recall
The affected units may contain extension sets with the microbore tubing instead of extension sets with the macrobore tubing.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
All consignees were notified by an Urgent: Product Recall Notification letter via Fed-Ex carrier on 04/02/2009. They were instructed not to use or further distribute any of the affected products. They were advised to check their inventory and to remove and return affected product to Bard Medical Division. A Recall & Effectiveness Check Form was included to be filled out and returned via fax, to Bard.
Quantity in commerce
16,100 units
Distribution
Nationwide Distribution - including states of AZ, CA, CO, GA, IA, MN, NE, and TX.
Product codes and lots
Lot #JUSJF441.