Z-1539-2009
Bard StatLock Intravenous (IV) Catheter Stabilization Kit, Catalog number: IVK0124, C.R. Bard, Inc...
C.R. Bard Inc / initiated 2009-04-02 / Terminated / root cause: manufacturing process
C.R. Bard, Inc., Urological Division began a recall of Bard StatLock Intravenous (IV) Catheter Stabilization Kit, Catalog number: IVK0124, C.R. Bard, Inc., Covington, GA 30014, USA on . The reason FDA records is "The affected units may contain extension sets with the microbore tubing instead of extension sets with the macrobore tubing". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1539-2009
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- C.R. Bard Inc
- Product
- Bard StatLock Intravenous (IV) Catheter Stabilization Kit, Catalog number: IVK0124, C.R. Bard, Inc., Covington, GA 30014, USA.
- Product code
- FPA Set, Administration, Intravascular
- Reason for recall
- The affected units may contain extension sets with the microbore tubing instead of extension sets with the macrobore tubing.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- All consignees were notified by an Urgent: Product Recall Notification letter via Fed-Ex carrier on 04/02/2009. They were instructed not to use or further distribute any of the affected products. They were advised to check their inventory and to remove and return affected product to Bard Medical Division. A Recall & Effectiveness Check Form was included to be filled out and returned via fax, to Bard.
- Quantity in commerce
- 16,100 units
- Distribution
- Nationwide Distribution - including states of AZ, CA, CO, GA, IA, MN, NE, and TX.
- Product codes and lots
- Lot #JUSJF441.