Z-0138-2010

Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System

GE Healthcare / initiated 2009-04-16 / Terminated / root cause: design

GE Medical Systems, LLC began a recall of Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System on . The reason FDA records is "Potential pump failure associated with the Coolix 4000 chiller of the Innova single plane and biplane systems that may stop the chiller from working. If a chiller failure occurs, the user will be informed with a message...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-0138-2010
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
GE Healthcare
Product
Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System.
Product code
JAA System, X-Ray, Fluoroscopic, Image-Intensified
Reason for recall
Potential pump failure associated with the Coolix 4000 chiller of the Innova single plane and biplane systems that may stop the chiller from working. If a chiller failure occurs, the user will be informed with a message on the n-room monitor stating that three minutes are left before the X-ray is inhibited. This could lead to a delay in treatment and possible additional X-ray exposure and contra
Root cause tag
design
FDA root cause
Device Design
Action
Consignees were sent on May 1, 2009 a GE Healthcare " Urgent Medical Device Correction" letter dated April 16, 2009. The letter was addressed to Hospital Administrator/ Risk Manger, Mangers of Radiology/ Cardiology, and Radiologists/ Cardiologists. The letter described the problem, Safety Issue, Affected Product Details, Safety Instructions, Product Correction and Contact Information.
Quantity in commerce
15 AND 46 UNLOCATED
Distribution
Worldwide distribution: USA, ARGENTINA, BELGIUM, BRAZIL, BULGARIA, CHINA, CROATIA, EGYPT, FRANCE, GERMANY, HONG KONG, HUNGARY, INDIA, ITALY, JAPAN, REPUBLIC OF KOREA, LATVIA, LIBYA, MEXICO, NORWAY, POLAND, RUSSIA, SINGAPORE, SPAIN, SWEDEN, TAIWAN, TURKEY, UNITED KINGDOM, ROMANIA, ISRAEL, UKRAINE, YEMEN, BELARUS, and AUSTRALIA.
Product codes and lots
SYSTEM ID:  0847160005 542120XR02  561548BIP  561881BP1  630856H3131 817465AR3131  985867INNOVA31 COR373301 EG1000VA01 LY1086VA01 M4166990  05448VAS01 VAD2121  YE1002VA01 YV0365   NOT LOCOLIZED S/N CHILLER 798080-21 796259-09 794637-08 795371-01 796133-12 796259-05 796464-08 796464-13 796464-15 796767-05 796941-03 796941-04 796941-07 796941-09 797178-05 797179-03 797179-08 797179-12 797449-03 797449-06 797632-02 797812-04 797812-09 797812-10 797812-12 767326-03 770355-02 772153-13 767746-05 785678-16 794637-11 799172-07 799347-03 799347-05 796260-01 783679-01 797178-03 797179-07 797179-09 797179-11 797812-14 797812-16 799347-02 798449-01 798449-02 798449-03