Z-1607-2009

Offset Adaptor Trial - 2MM, Triathlon Revision Instruments. Non-Sterile; Howmedica Osteonics Corp: 325...

Howmedica Osteonics Corp / initiated 2009-04-16 / Terminated / root cause: manufacturing process

Stryker Howmedica Osteonics Corp. began a recall of Offset Adaptor Trial - 2MM, Triathlon Revision Instruments. Non-Sterile; Howmedica Osteonics Corp: 325 Corporate Drive, Mahwah, NJ 07430. Indications for use... on . The reason FDA records is "Triathlon Offset Adaptors may seize during surgery and the OR staff may be unable to disassemble the instrument using the removal tool". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1607-2009
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Howmedica Osteonics Corp
Product
Offset Adaptor Trial - 2MM, Triathlon Revision Instruments. Non-Sterile; Howmedica Osteonics Corp: 325 Corporate Drive, Mahwah, NJ 07430. Indications for use include disabling joint disease of the knee.
Product code
JWH Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer
Reason for recall
Triathlon Offset Adaptors may seize during surgery and the OR staff may be unable to disassemble the instrument using the removal tool.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
"Urgent Product Correction" letters dated April 16, 2009 were sent to all Stryker Representatives, Hospital Rick Management Departments, Hospital Chief of Orthopaedics, Surgeon and countries by Federal Express. The notice explains the Issue, Potential Hazards and Risk Mitigation Factors of the affected product(s) including a request to complete and return the Product Correction Form. For further questions, contact Stryker Orthopedics, Division of Regulatory Reporting at 201-831-5970.
Quantity in commerce
2498 units total, all sizes.
Distribution
Worldwide Distribution.
Product codes and lots
Catalog Number 5570-T-020; All Lot Numbers.