Z-1686-2009

DePuy P.F.C. Sigma Knee System, Non-Porous Cruciate Retaining Femoral Component, 73 mm M/L 69 mm A/P, 5 left...

Depuy Orthopaedics Inc. / initiated 2009-05-07 / Terminated / root cause: manufacturing process

Depuy Orthopaedics, Inc. began a recall of DePuy P.F.C. Sigma Knee System, Non-Porous Cruciate Retaining Femoral Component, 73 mm M/L 69 mm A/P, 5 left, sterile, DePuy Orthopaedics, Inc., Warsaw, IN... on . The reason FDA records is "There may be a crack on the lateral side of the condyle in the posterior chamfer region. The device is used as a Orthopedic knee implant". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1686-2009
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Depuy Orthopaedics Inc.
Product
DePuy P.F.C. Sigma Knee System, Non-Porous Cruciate Retaining Femoral Component, 73 mm M/L 69 mm A/P, 5 left, sterile, DePuy Orthopaedics, Inc., Warsaw, IN; REF 96-0005. The device is used for total or unicompartmental knee arthroplasty and is intended to provide increased patient mobility and reduced pain by replacing the damaged knee joint articulation in patients where there is evidence of sufficient sound bone to seat and support the components.
Product code
JWH Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer
Reason for recall
There may be a crack on the lateral side of the condyle in the posterior chamfer region. The device is used as a Orthopedic knee implant.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
The firm's sales staff was notified by email on May 7, 2009, and the recall was expanded to include additional lots on May 29, 2009, instructing them to remove the affected lots from their inventory and the inventory of any consignees, to provide the consignees with a copy of the recall letter, and to return the product to the firm. Hospital consignees were notified by letter dated May 7, 2009, as well as by updated letter on May 29, 2009. A dear doctor letter dated May 29, 2009 was issued to implanting surgeons advising them of the issue. Direct questions to Depuy Customer Service by calling 1-800-366-8143.
Quantity in commerce
260
Distribution
Nationwide.
Product codes and lots
Lot Numbers: 2833275, 2833276, 2833277, 2833278, 2833279, 2835157, 2835908, 2835909, 2837020, 2838165, 2839666, 2839667, 2839668, 2839670, 2839671, 2839672, 2839675, 2840479, 2840480, 2841253, 2842526, 2842527, 2842528, 2842930, 2843263, 2843265, 2843267, 2843268, 2843270, 2844381, 2844383, 2844386, 2844796, 2844799, 2844800, 2845530, 2845534, 2846326, 2846330, 2846332, 2852392, 2852394, 2855015, 2855016, 2857271, 2857272, 2857881, 2857882, 2857884, 2857885, 2858572, 2858573, 2858574, 2858576, 2858577, 2859195, 2859197, 2859198, 2859199, 2859915, 2859916, 2859917, 2859920, 2860448, 2860449, 2860452, 2860455, 2860456, 2861519, 2861520, 2861699, 2862589, 2862591, 2863841, 2863843, 2864833, 2864834, 2864835, 2864837, 2864839, 2866038, 2868313, 2868314, 2869119, 2869120, 2869121, 2869122, 2869123, 2871182, 2871183, 2871835, 2871836, 2874434, 2874435, 2874436, 2876307, 2876309, 2876311, 2876314, 2881638, 2885685, 2885689, 2886278, 2886968 and 2886974.