Z-1676-2009

Symbia S Series SPECT System; Single-Photon Emission Computed Tomography; Siemens Medical Solutions USA...

Siemens Medical Solutions USA, Inc. / initiated 2009-04-22 / Terminated

Siemens Medical Solutions USA, Inc. began a recall of Symbia S Series SPECT System; Single-Photon Emission Computed Tomography; Siemens Medical Solutions USA, Inc., Hoffman Estates, IL 60195-2061 Indications for... on . The reason FDA records is "The pinion gear could crack and cause the gantry to spin freely thereby damaging the Symbia S or T System and potentially injuring a patient during a scan". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-1676-2009
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Medical Solutions USA, Inc.
Product
Symbia S Series SPECT System; Single-Photon Emission Computed Tomography; Siemens Medical Solutions USA, Inc., Hoffman Estates, IL 60195-2061 Indications for use: SPECT + CT: Perform scans and nuclear imaging studies with the same instrument. To obtain attenuation corrected images and to provide registration of anatomical and physiological images within the patient's anatomy.
Product code
KPS System, Tomography, Computed, Emission
Reason for recall
The pinion gear could crack and cause the gantry to spin freely thereby damaging the Symbia S or T System and potentially injuring a patient during a scan.
FDA root cause
Finished device change control
Action
Siemens sent Urgent Field Correction Recall letters dated 4/22/09 to the affected users of the Symbia S and Symbia T Systems. The accounts were informed of the potential faulty pinion gear which could crack and cause the gantry to spin freely, damaging the system and potentially injuring a patient during a scan. The customers were advised to stop using the system if they hear any abnormal grinding noise from the gantry. They also were advised that a Siemens Customer Service Representative would contact them within 60 days to schedule a replacement of the pinion gear free of charge. Any questions were directed to contact Siemens Medical Solutions USA, Inc. at 1-800-767-2313 (USA) for assistance. The accounts were requested to complete the Mandatory Fax-Back Form acknowledging receipt and understanding of the letter.
Quantity in commerce
37 units
Distribution
Worldwide Distribution -- USA, Australia, Austria, Brazil, Bulgaria, Canada, China, Finland, France, Germany, India, Italy, Japan, Mexico, Netherlands, Norway, Philippines, Russia, Spain, Taiwan, and Turkey.
Product codes and lots
part 8717741, serial numbers 1326, 1315, 1324, 1330, 1310, 1325, 1336, 1307, 1317, 1331, 1299, 1300, 1301, 1302, 1303, 1304, 1305, 1306, 1308, 1309, 1311, 1312, 1313, 1314, 1316, 1318, 1319, 1320, 1322, 1323, 1327, 1328, 1329, 1332, 1334, 1335, and 1337.