Z-1525-2009
Stryker Xia Dual Anterior Staple; Non Sterile; Catalog number 03820115; Manufactured by Stryker Spine SAS...
Howmedica Osteonics Corp / initiated 2007-04-12 / Terminated / root cause: labeling and UDI
Stryker Spine began a recall of Stryker Xia Dual Anterior Staple; Non Sterile; Catalog number 03820115; Manufactured by Stryker Spine SAS, Cestas, France; Distributed by Howmedica Osteonics... on . The reason FDA records is "Labeling Issue: The large Xia Anterior Staple was laser marked with an "S", indicating small". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.
Recall record
- Recall number
- Z-1525-2009
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Howmedica Osteonics Corp
- Product
- Stryker Xia Dual Anterior Staple; Non Sterile; Catalog number 03820115; Manufactured by Stryker Spine SAS, Cestas, France; Distributed by Howmedica Osteonics Corp, Mahwah, NJ 07430 Single use device sterilized prior to use. For anterior/anterolateral and posterior noncervical pedicle and non-pedicle fixation for the following: degenerative disc disease, spondylolisthesis, trauma, spinal stenosis, curvatures, tumor, psudoartrosis, and failed previous fusion.
- Product code
- NKB Thoracolumbosacral Pedicle Screw System
- Reason for recall
- Labeling Issue: The large Xia Anterior Staple was laser marked with an "S", indicating small.
- Root cause tag
- labeling and UDI
- FDA root cause
- Error in labeling
- Action
- Market Withdrawal letters were sent to Stryker branches who received product on April 13, 2007 by Federal Express. The letters asked that they examine their inventory and hospital locations to identify the implant; reconcile these implants on the attached Product Accountability Form and fax a copy to (201) 760-8456; retrieve and return the affected implants; and contact their Stryker Customer Service rep to re-order the product that they are returning as part of the regulatory action.
- Quantity in commerce
- 38 in US
- Distribution
- Nationwide Distribution.
- Product codes and lots
- lot code HJ3