Z-1487-2009

Boston Scientific, Fort Guide Wire, Floppy, 300cm x 2 cm REF 34949-04, for use in intravascular...

Boston Scientific Corporation / initiated 2009-05-12 / Terminated / root cause: manufacturing process

Boston Scientific Corporation began a recall of Boston Scientific, Fort Guide Wire, Floppy, 300cm x 2 cm REF 34949-04, for use in intravascular interventional procedures on . The reason FDA records is "Product packaged incorrectly. One batch/lot of wires, that should contain radiopaque marker bands, is missing the marker bands; and a second batch/lot of wires, which should not contain radiopaque marker bands, has the...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1487-2009
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Boston Scientific Corporation
Product
Boston Scientific, Fort Guide Wire, Floppy, 300cm x 2 cm REF 34949-04, for use in intravascular interventional procedures.
Product code
DQX Wire, Guide, Catheter
Reason for recall
Product packaged incorrectly. One batch/lot of wires, that should contain radiopaque marker bands, is missing the marker bands; and a second batch/lot of wires, which should not contain radiopaque marker bands, has the marker bands. Since the marker bands issue would not be visually apparent to the physician there is a potential for prolongation or delay of the procedure in order to exchange the
Root cause tag
manufacturing process
FDA root cause
Process control
Action
Consignees were sent a "Boston Scientific Urgent Medical Device Recall" letter dated may 19, 2009. The letter was addressed to "Dear Risk Manger/ Field Action Contact". The letter described the problem, customer steps for recall and listing of product batches / lots. The letter requested consignees to return the Reply Verification Tracking Form.
Quantity in commerce
Total of 4 boxes (5 units /box).
Distribution
USA distribution: AR, MO, CA, FL, and NV.
Product codes and lots
01403876