Philips Medical Systems North America Co. Phillips began a recall of Product's marketing brochure is labeled in part: "Philips Easy Diagnost Eleva Conventional R/F system...PHILIPS". manufactured by Philips Medical Systems Gmbh... on . The reason FDA records is "Patient and image data can be exchanged when a patient is selected in the patient list and the examination tab is pressed with background query running at the same time. There is a risk that the data may then be linked...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-0479-2011
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Philips North America LLC
- Product
- Product's marketing brochure is labeled in part: "Philips Easy Diagnost Eleva Conventional R/F system...PHILIPS". manufactured by Philips Medical Systems Gmbh, Hamburg, Germany.
- Product code
- KPR System, X-Ray, Stationary
- Reason for recall
- Patient and image data can be exchanged when a patient is selected in the patient list and the examination tab is pressed with background query running at the same time. There is a risk that the data may then be linked to the incorrect patient, which may adversely affect diagnosis and treatment.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- Philips Medical Systems issued URGENT-Device Correction letters dated January 27, 2009 to consignees describing the issue with the Easy Diagnost Eleva GXR RF systems. The consignees were informed of the potential for incorrect diagnosis or treatment if an incorrect image is used for diagnosis. Consignees should contact the Philips Healthcare Call Center at 800-722-9377, #5, #2 and reference "FC0 70600030" with any communication.
- Quantity in commerce
- 314 units
- Distribution
- Nationside Distribution: USA, including the states of AK, AL, AR, AZ, CA, CO, CT, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, NC, ND, NE, NE, NH, NJ, NV, NY, OH, OK, OR, PA, PA, RI, SC, SD, TN, TX, TX, UT, VA, VT, WA, WI, and WV. Devices are not distributed to any foreign consignees.
- Product codes and lots
- Site Numbers: 506250, 506251, 506276, 519152, 520209, 521153, 530573, 530607, 532426, 532779, 533491, 533574, 534207, 534419, 534644, 534647, 534675, 534803, 534945, 535378, 535437, 535785, 536051, 536458, 536522, 536636, 536637, 536687, 536765, 536840, 536933, 536941, 536973, 536985, 536994, 537013, 537087, 537091, 537179, 537186, 537492, 537518, 537571, 537689, 537761, 537796, 537803, 537863, 537945, 537946, 537970, 537971, 538237, 538328, 538351, 538407, 538451, 538458, 538462, 538477, 538480, 538519, 538759, 538812, 538814, 538866, 538891, 538901, 539040, 539095, 539101, 539103, 539237, 539291, 539310, 539339, 539377, 539378, 539379, 539411, 539419, 539488, 539501, 539519, 539527, 539529, 539530, 539557, 539590, 539683, 539785, 540022, 540163, 540187, 540361, 540380, 540445, 540457, 540459, 540624, 540788, 540810, 541011, 541054, 541063, 541078, 541083, 541093, 541348, 541402, 541494, 541535, 541614, 541694, 541721, 541803, 541838, 541850, 541872, 541884, 542061, 542121, 542122, 542205, 542308, 542401, 542521, 542537, 542589, 542593, 542716, 542829, 542849, 543136, 543144, 543224, 543241, 543249, 543275, 543297, 543331, 543369, 543415, 543426, 543619, 543620, 543660, 543723, 543754, 543755, 543771, 543776, 543803, 543841, 544075, 544113, 544172, 544228, 544236, 544297, 544341, 544417, 544480, 544481, 544568, 544648, 544654, 544695, 544937, 544999, 545000, 545038, 545120, 545163, 545166, 545213, 545231, 545234, 545244, 545391, 545452, 545710, 545746, 545780, 545783, 545824, 545844, 545923, 545948, 545951, 546029, 546155, 546490, 546524, 546551, 546583, 546602, 547054, 547189, 547243, 547297, 547304, 547371, 547598, 547599, 547655, 547657, 547744, 547758, 547759, 547798, 547936, 547937, 548021, 548276, 548290, 548302, 548382, 548559, 548784, 549008, 549080, 549269, 549270, 549381, 549416, 549509, 549557, 549616, 549663, 550072, 550079, 550197, 550313, 550591, 550804, 550936, 550960, 551005, 551416, 551573, 551616, 551645, 551721, 551759, 551941, 551954, 551980, 552204, 552205, 552211, 552324, 552363, 552364, 552407, 552502, 552579, 552594, 552626, 552640, 552917, 552918, 552965, 552966, 553013, 553172, 553207, 553224, 553230, 553275, 553349, 553387, 553441, 553448, 553455, 553563, 553850, 553893, 553900, 553926, 554061, 554483, 554581, 554641, 554721, 554722, 554769, 555052, 555269, 555458, 555524, 555574, 555626, 555964, 555993, 556124, 556621, 556622, 556726, 556785, 556984, 557177, 557259, 557272, 558437, 558596, 558734, 558775, 41443799, 41444441, 41658234, 41862667, 42039161, and 537967 / 49371595.