BD Diagnostics, BD GeneOhm MRSA Assay, REF: 441242 (200 Reaction Kit) and 441244 (48 Reaction Kit). For the...
BD Diagnostics (GeneOhm Sciences, Inc) began a recall of BD Diagnostics, BD GeneOhm MRSA Assay, REF: 441242 (200 Reaction Kit) and 441244 (48 Reaction Kit). For the rapid detection of methicillin resistant MRSA DNA... on . The reason FDA records is "BD Diagnostics GeneOhm (BDDG) has determined that the Cepheid SmartCycler reaction tubes packaged in the BD GeneOhm MRSA test kits identified above may be defective". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.
Recall record
- Recall number
- Z-1963-2009
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- BD Diagnostics (Geneohm Sciences, Inc.)
- Product
- BD Diagnostics, BD GeneOhm MRSA Assay, REF: 441242 (200 Reaction Kit) and 441244 (48 Reaction Kit). For the rapid detection of methicillin resistant MRSA DNA in nasal specimen.
- Product code
- NQX System, Nucleic Acid Amplification Test, Dna, Methicillin Resistant Staphylococcus Aureus, Direct Specimen
- Reason for recall
- BD Diagnostics GeneOhm (BDDG) has determined that the Cepheid SmartCycler reaction tubes packaged in the BD GeneOhm MRSA test kits identified above may be defective.
- Root cause tag
- packaging
- FDA root cause
- Other
- Action
- BD Diagnostics GeneOhm began contacting consignees by phone November 17, 2008 followed by a letter dated November 21, 2008 instructing discontinued use of the affected lots. Consignees were also asked to complete and return the enclosed Field Action Verification form. To arrange for product replacement or for further information, contact BD Diagnostics Technical Service at 1-888-436-3646 extension 2.
- Quantity in commerce
- 114 kits
- Distribution
- Worldwide Distribution -- US (CA, IL, IN, KT, MD, ME, MN, MO, MT, NC, NJ, NY, OK, PA, TN, TX, UT, VA, and WI), Belgium and Japan.
- Product codes and lots
- Catalog Number: 441242, Lot Numbers: 03T08206, 03T08207, 03T08207a and 03T08221a; and Catalog Number: 441244, Lot Number: 03T08218.