Medtronic INTREPID Spinal System SPACER L 42X30, 12 DEG, Size: 10MM, Part Number: REF 7963210; 12MM, REF...
Medtronic Sofamor Danek USA Inc began a recall of Medtronic INTREPID Spinal System SPACER L 42X30, 12 DEG, Size: 10MM, Part Number: REF 7963210; 12MM, REF 7963212, 14MM, REF 7963214; 16MM, REF 7963216; 18MM... on . The reason FDA records is "The firm received complaints regarding the use of the product in patients with poor bone quality, translational instability, difficulty implanting the cover plate and case of cover plate detachment". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-1801-2009
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Medtronic Sofamor Danek USA Inc
- Product
- Medtronic INTREPID Spinal System SPACER L 42X30, 12 DEG, Size: 10MM, Part Number: REF 7963210; 12MM, REF 7963212, 14MM, REF 7963214; 16MM, REF 7963216; 18MM, REF 7963218; 20MM, REF 7963220; USA, Rx only, STERILE R. The INTREPID Spinal System is indicated for use with autogenous bone graft in patients with degenerative disc disease.
- Product code
- MAX Intervertebral Fusion Device With Bone Graft, Lumbar
- Reason for recall
- The firm received complaints regarding the use of the product in patients with poor bone quality, translational instability, difficulty implanting the cover plate and case of cover plate detachment.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- Medtronic Sofamor Danek USA, Inc. instructed Sales Representative on starting February 6, 2009 by phone to immediately return product from the field. The firm further issued a "Subject: Medtronic Intrepid Intervertebral Body Fusion Device Recall" notice dated April 2009. The recall notices were mailed directly to Risk Managers and Surgeons, with Recall Questionnaires and Surgeon Confirmation forms included respectively. For further questions, contact your Medtronic Sales Representative or the Medtronic Global Quality Department at 1-800-876-3133 extension 6333.
- Quantity in commerce
- 345 units
- Distribution
- Worldwide Distribution -- United States, Australia, Germany, The Netherlands, and South Africa.
- Product codes and lots
- All Lots.