Z-1594-2009

Roche Coaguchex XS Systems for Patient Self-Testing, Roche Diagnostics, Indianapolis, IN; REF 04837738001...

Roche Diagnostics GmbH / initiated 2009-05-13 / Terminated

Roche Diagnostics Corp. began a recall of Roche Coaguchex XS Systems for Patient Self-Testing, Roche Diagnostics, Indianapolis, IN; REF 04837738001 includes the Roche ACCU-CHEK Softclix Lancing Device... on . The reason FDA records is "The lancet cap may be missing, which could result in an unintended lancet stick to the user". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-1594-2009
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Roche Diagnostics GmbH
Product
Roche Coaguchex XS Systems for Patient Self-Testing, Roche Diagnostics, Indianapolis, IN; REF 04837738001 includes the Roche ACCU-CHEK Softclix Lancing Device, Distributed by Roche Diagnostics, Indianapolis, IN. Model 3144844001 (10 lancets). The device is used to prick the finger and to obtain a drop of blood in preparation for diabetes/insulin level testing.
Product code
GJS Test, Time, Prothrombin
Reason for recall
The lancet cap may be missing, which could result in an unintended lancet stick to the user.
FDA root cause
Other
Action
An "Urgent Medical Device Correction" letter dated May 13, 2009 was sent to all affected customers, distributors and consumer consignees who have either registered their blood glucose meter, contacted the firm about the lancets since 10/1/08, or have been shipped lancets by Roche since 10/1/08 were notified of the problem. Distributors were requested to notify each patient self-tester to whom a meter kit was distributed since October 1, 2008. The firm issued a press release on May 18, 2009. Diabetes Care customers: If you have questions regarding the recall, please contact the ACCU-CHEK Softclix Lancet Hotline at 1-800-778-7057. CoaguChek meter customers: If you have questions regarding the recall, please contact the ACCU-CHEK Softclix Lancet Hotline at 1-800-778-7505.
Quantity in commerce
449,533 packages of lancets
Distribution
Nationwide.
Product codes and lots
REF 04837738001; 117200, 117202, 117203, 117204, 117205, 117208, 117209, 117213, 117216, 117219, 117220, 117222, 117227, 117228, 117229, 117231, 117233, 117235, 117237, 117238, 117240 and 117241. Lancet Model 3144844001; lots M25C2, M25C5, M25C7, N32A5 and M32B5.