Stryker 550 mL AutoFuser PainPump, dual site continuous, 4 mL/hr, 2 x 10.0 in. catheters, sterile, REF...
Stryker Instruments Div. of Stryker Corporation began a recall of Stryker 550 mL AutoFuser PainPump, dual site continuous, 4 mL/hr, 2 x 10.0 in. catheters, sterile, REF 0535-400-210 on . The reason FDA records is "Failure to infuse and potential compromised sterility: The seal on the tyvek tray cover may fail, compromising sterility, and the device may fail to infuse the drug, resulting in a loss of pain mitigation". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1853-2009
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Stryker Corp.
- Product
- Stryker 550 mL AutoFuser PainPump, dual site continuous, 4 mL/hr, 2 x 10.0 in. catheters, sterile, REF 0535-400-210.
- Product code
- MEB Pump, Infusion, Elastomeric
- Reason for recall
- Failure to infuse and potential compromised sterility: The seal on the tyvek tray cover may fail, compromising sterility, and the device may fail to infuse the drug, resulting in a loss of pain mitigation.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- Stryker sent all consignees a Medical Device Recall Notification letter dated 5/14/09 and instructed to locate and destroy the recalled products and to complete and return the business reply card.
- Quantity in commerce
- 28,755 of all pumps
- Distribution
- Nationwide and Canada.
- Product codes and lots
- Lots 08142756254 through 09125933719.