Z-1616-2009

Zimmer Alumina Ceramic Femoral Head, 12/14 taper, 32 mm diameter, neck length 0 mm, sterile, Zimmer, Warsaw...

Zimmer GmbH / initiated 2009-05-29 / Terminated / root cause: packaging

Zimmer Inc. began a recall of Zimmer Alumina Ceramic Femoral Head, 12/14 taper, 32 mm diameter, neck length 0 mm, sterile, Zimmer, Warsaw, IN; Catalog Number: . Orthopedic implant... on . The reason FDA records is "The corner tab of the inner compartment lid was not folded under before the outer lid was sealed, resulting in lack of assurance of sterility". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.

Recall record

Recall number
Z-1616-2009
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Zimmer GmbH
Product
Zimmer Alumina Ceramic Femoral Head, 12/14 taper, 32 mm diameter, neck length 0 mm, sterile, Zimmer, Warsaw, IN; Catalog Number: 6428-32-02. Orthopedic implant used in total hip arthroplasty.
Product code
LZO Prosthesis, Hip, Semi-Constrained, Metal/Ceramic/Polymer, Cemented Or Non-Porous, Uncemented
Reason for recall
The corner tab of the inner compartment lid was not folded under before the outer lid was sealed, resulting in lack of assurance of sterility.
Root cause tag
packaging
FDA root cause
Packaging process control
Action
Zimmer, Inc. issued "Urgent: Device Recall" notices dated May 29, 2009 informing Consignees of the affected devices. Users were instructed to locate and return affected product and complete and return an Inventory Return Certification form via fax to the firm at 1-547-372-4265. For further questions, contact Zimmer, Inc. at 1-800-613-6131.
Quantity in commerce
147 of both products
Distribution
Worldwide Distribution -- United States, Brazil, Canada, Chile, Czech Republic, Finland, France, Germany, Italy, Korea, Lebanon, Switzerland and United Kingdom.
Product codes and lots
Lot Number: 61007046.