Zimmer Kinectivo Technology Modular Neck, BB, 12/14 neck taper, sterile, Zimmer, Warsaw, IN; Catalog Number...
Zimmer Inc. began a recall of Zimmer Kinectivo Technology Modular Neck, BB, 12/14 neck taper, sterile, Zimmer, Warsaw, IN; Catalog Number: 00-7848-022-01. Orthopedic implant used in total... on . The reason FDA records is "The corner tab of the inner compartment lid was not folded under before the outer lid was sealed, resulting in lack of assurance of sterility". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.
Recall record
- Recall number
- Z-1617-2009
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Zimmer GmbH
- Product
- Zimmer Kinectivo Technology Modular Neck, BB, 12/14 neck taper, sterile, Zimmer, Warsaw, IN; Catalog Number: 00-7848-022-01. Orthopedic implant used in total hip arthroplasty.
- Product code
- LPH Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Porous Uncemented
- Reason for recall
- The corner tab of the inner compartment lid was not folded under before the outer lid was sealed, resulting in lack of assurance of sterility.
- Root cause tag
- packaging
- FDA root cause
- Packaging process control
- Action
- Zimmer, Inc. issued "Urgent: Device Recall" notices dated May 29, 2009 informing Consignees of the affected devices. Users were instructed to locate and return affected product and complete and return an Inventory Return Certification form via fax to the firm at 1-547-372-4265. For further questions, contact Zimmer, Inc. at 1-800-613-6131.
- Quantity in commerce
- 147 of both products.
- Distribution
- Worldwide Distribution -- United States, Brazil, Canada, Chile, Czech Republic, Finland, France, Germany, Italy, Korea, Lebanon, Switzerland and United Kingdom.
- Product codes and lots
- Lot Number: 60917793.