Inovel N95 HealthCare Particulate Respirator, Model/Catalog Number: FRN95-SEZ, Expiration Date: 07/13/2010

Moldex/Metric, Inc. / initiated 2008-07-23 / Terminated / root cause: packaging

Inovel LLC began a recall of Inovel N95 HealthCare Particulate Respirator, Model/Catalog Number: FRN95-SEZ, Expiration Date: 07/13/2010 on . The reason FDA records is "Display boxes labeled as FRN95-MLEZ, Medium/Large Size Respirators were inadvertently mixed with Display Boxes labeled for FRN95-SEZ Small Size Respirators". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.

Recall record

Recall number
Z-1576-2009
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Moldex/Metric, Inc.
Product
Inovel N95 HealthCare Particulate Respirator, Model/Catalog Number: FRN95-SEZ, Expiration Date: 07/13/2010.
Product code
MSH Respirator, Surgical
Reason for recall
Display boxes labeled as FRN95-MLEZ, Medium/Large Size Respirators were inadvertently mixed with Display Boxes labeled for FRN95-SEZ Small Size Respirators.
Root cause tag
packaging
FDA root cause
Packaging process control
Action
The recalling process was initiated on July 23, 2007. The affected distributor was notified of the recall through a certified recall letter with an attached Customer Response card. The distributor was instructed to notify and obtain recall product from their customers and return all affected product to Inovel. Direct questions about the recall to Inovel at 1-866-546-6835.
Quantity in commerce
80 cases
Distribution
US (Illinois only).
Product codes and lots
Lot Number: 060713-244.