Z-1320-2010
ThermoFisher Scientific Microm STP 420D automated tissue processor
Thermo Fisher Scientific Oy / initiated 2009-06-03 / Terminated
Thermo Fisher Scientific began a recall of ThermoFisher Scientific Microm STP 420D automated tissue processor on . The reason FDA records is "The firm has received complaints regarding specimen integrity (higher rate of specimen fragmentation) as well as complaints that the chamber may stop rotating prematurely, resulting in sample deterioration". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-1320-2010
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Thermo Fisher Scientific Oy
- Product
- ThermoFisher Scientific Microm STP 420D automated tissue processor.
- Product code
- IEO Processor, Tissue, Automated
- Reason for recall
- The firm has received complaints regarding specimen integrity (higher rate of specimen fragmentation) as well as complaints that the chamber may stop rotating prematurely, resulting in sample deterioration.
- FDA root cause
- Other
- Action
- Consignees were notified of the problem by letter, on or about 6/3/09, updating the instructions for use. The units were subsequently upgraded on or before 9/8/09. For additional questions and information, please contact ThermoFisher's Technical Support department , Subject: PT Module Advisory Notice.
- Quantity in commerce
- 16
- Distribution
- Hawaii, Kentucky, Louisiana, Michigan, Missouri, Nebraska, New York, North Carolina, Ohio, Pennsylvania, Tennessee, Texas and West Virginia.
- Product codes and lots
- All units.