Z-1578-2009

TumorLoc software application, Software Versions: 23.1.1, 20.7.13, and 20.8.15 for use on the GEMINI TF 16...

Philips North America LLC / initiated 2009-02-19 / Terminated / root cause: software

Philips Medical Systems (Cleveland) Inc began a recall of TumorLoc software application, Software Versions: 23.1.1, 20.7.13, and 20.8.15 for use on the GEMINI TF 16 and GEMINI TF 64 Slice PET/CT Systems, Catalog... on . The reason FDA records is "A software anomaly in the TumorLoc software for the firm's Brilliance CT scanners was discovered. The same software anomaly was subsequently determined to be present in the firm's GEMINI PET/CT units. The software...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-1578-2009
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Philips North America LLC
Product
TumorLoc software application, Software Versions: 23.1.1, 20.7.13, and 20.8.15 for use on the GEMINI TF 16 and GEMINI TF 64 Slice PET/CT Systems, Catalog Numbers: 882470 - GEMINI TF 16 Slice; 882471 - GEMINI TF 64 Slice; and 882473 - GEMINI TF 16 Slice Base Configuration. Model Numbers: 4535 679 83931 - GEMINI TF 16 Slice; 4535 679 94741 - GEMINI TF-64 Slice and 4535 674 41711 - GEMINI TF 16 Slice Base Configuration. Tumor Localization is an application used in radiation oncology departments for localizing target volumes for radiation therapy planning. It includes exclusive features for viewing Respiratory Correlated CT data sets and analyzing motion of target and surrounding anatomy.
Product code
JAK System, X-Ray, Tomography, Computed
Reason for recall
A software anomaly in the TumorLoc software for the firm's Brilliance CT scanners was discovered. The same software anomaly was subsequently determined to be present in the firm's GEMINI PET/CT units. The software defect results in a the potential for data to be lost during reconstruction of images in which slices can either be discarded, or interleaved based upon their series ID.
Root cause tag
software
FDA root cause
Software design
Action
An "URGENT - Field Safety Notice" dated February 12, 2009 was sent via FedEx to domestic customers. The notice described the affected products, issue, involved hazard should problem occur and corrective actions for the customer/user. The letter also notified users/customers that a Philips Healthcare Field Service Engineer will contact and visit their site within 6 months to install a TumorLoc upgrade. Notification will also be sent to the international consignees by Philips personnel. For further information or support concerning this issue, please contact your local Philips representative: Customer Care Center (1-800-722-9377, option 5: Diagnostic Imaging, option 5: Radiation Therapy, option 3: Oncology) or to your local Philips Healthcare office.
Quantity in commerce
13 units
Distribution
Worldwide Distribution -- US, Columbia, Germany, France, Netherlands, Canada, Belgium and Slovenia.
Product codes and lots
GEMINI TF 16 Slice, Serial Numbers: 7035, 7052, 7093, 7096, 7123, 7128, and 7133; GEMINI TF 16 Slice base configuration, Serial Number: 7150; and GEMINI TF 64 Slice, Serial Numbers: 7007, 7088, 7103, 7126, and 7132.