Z-1704-2009
IMHS CP, 3.2mm, Tip Threaded Guide Pin, REF71687000, for use with IntraMedullary Hip Screw in orthopedic...
Smith & Nephew, Inc. / initiated 2009-01-22 / Terminated / root cause: manufacturing process
Smith & Nephew Inc began a recall of IMHS CP, 3.2mm, Tip Threaded Guide Pin, REF71687000, for use with IntraMedullary Hip Screw in orthopedic surgery on . The reason FDA records is "Incorrect depth gauge mark etched onto guide wire pin. Guide pin is too long". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1704-2009
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Smith & Nephew, Inc.
- Product
- IMHS CP, 3.2mm, Tip Threaded Guide Pin, REF71687000, for use with IntraMedullary Hip Screw in orthopedic surgery.
- Product code
- LXH Orthopedic Manual Surgical Instrument
- Reason for recall
- Incorrect depth gauge mark etched onto guide wire pin. Guide pin is too long.
- Root cause tag
- manufacturing process
- FDA root cause
- Process design
- Action
- Urgent Medical Device Recall letter dated January 23, 2009. All affected Smith & Nephew Sales Representatives, Distributors and Hospitals were notified of problem and the recall via Fed Ex and e-mail on 01/22/2009. They were instructed to immediately check their inventory and return the product.
- Quantity in commerce
- 99 units
- Distribution
- Distribution to USA, Switzerland and Australia.
- Product codes and lots
- Lot Number: 08K22270