TumorLoc Software Application for the EBW (Extended Brilliance Workstation, Model # 455011002031, sold as an...

Philips North America LLC / initiated 2009-02-19 / Terminated / root cause: software

Philips Medical Systems (Cleveland) Inc began a recall of TumorLoc Software Application for the EBW (Extended Brilliance Workstation, Model # 455011002031, sold as an accessory for the GEMINI PET/CT Systems running CT... on . The reason FDA records is "A software anomaly in the TumorLoc software for the firm's Brilliance CT scanners was discovered. The same software anomaly was subsequently determined to be present in the firm's GEMINI PET/CT units. The software...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-1580-2009
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Philips North America LLC
Product
TumorLoc Software Application for the EBW (Extended Brilliance Workstation, Model # 455011002031, sold as an accessory for the GEMINI PET/CT Systems running CT software versions 3.5.x. , CT host 2.2. Tumor Localization is an application used in radiation oncology departments for localizing target volumes for radiation therapy planning. It includes exclusive features for viewing Respiratory Correlated CT data sets and analyzing motion of target and surrounding anatomy.
Product code
KPS System, Tomography, Computed, Emission
Reason for recall
A software anomaly in the TumorLoc software for the firm's Brilliance CT scanners was discovered. The same software anomaly was subsequently determined to be present in the firm's GEMINI PET/CT units. The software defect results in a the potential for data to be lost during reconstruction of images in which slices can either be discarded, or interleaved based upon their series ID.
Root cause tag
software
FDA root cause
Software design
Action
An "URGENT - Field Safety Notice" dated February 12, 2009 was sent via FedEx to domestic customers. The notice described the affected products, issue, involved hazard should problem occur and corrective actions for the customer/user. The letter also notified users/customers that a Philips Healthcare Field Service Engineer will contact and visit their site within 6 months to install a TumorLoc upgrade. Notification will also be sent to the international consignees by Philips personnel. For further information or support concerning this issue, please contact your local Philips representative: Customer Care Center (1-800-722-9377, option 5: Diagnostic Imaging, option 5: Radiation Therapy, option 3: Oncology) or to your local Philips Healthcare office.
Quantity in commerce
11 units
Distribution
Worldwide Distribution -- US, Columbia, Germany, France, Netherlands, Canada, Belgium and Slovenia.
Product codes and lots
Serial Numbers: 7909, 8940, 8804, 9019, 10331, 10348, 10701, 11555, 55110, 55172 and 55180.