Boston Scientific EndoVive 20Fr (6.7 mm) Safety PEG Kit, Push Method, latex free, sterile, Boston Scientific...
Boston Scientific Corp began a recall of Boston Scientific EndoVive 20Fr (6.7 mm) Safety PEG Kit, Push Method, latex free, sterile, Boston Scientific Corp., Natick, MA; Cat. no./REF 6653, inner pouch... on . The reason FDA records is "The user may be unable to pass the device over the guidewire for placement in the patient, resulting in an inability to use the device". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1627-2009
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Boston Scientific Corporation
- Product
- Boston Scientific EndoVive 20Fr (6.7 mm) Safety PEG Kit, Push Method, latex free, sterile, Boston Scientific Corp., Natick, MA; Cat. no./REF 6653, inner pouch UPN M00566530 and outer carton UPN M00566531. The devices are indicated for providing nutrition directly into the stomach in adult and pediatric patients who are unable to consume nutrition by conventional means.
- Product code
- KNT Tubes, Gastrointestinal (And Accessories)
- Reason for recall
- The user may be unable to pass the device over the guidewire for placement in the patient, resulting in an inability to use the device.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- Boston Scientific Corporation issued an "Urgent Medical Device Recall - Immediate Action Required" notice dated May 5, 2009 as follow-up to their "Recall of the Safety and Standard Push PEG Kits" dated December 30, 2008. The notice outlined the issue and Consignees were instructed to locate and return all affected product to the firm. In addition, the firm requests that users complete and return a Reply Verification Tracking Form via fax to 1-508-683-5578. For further questions, contact the Boston Scientific Corporation Field Action Team by phone at 1-508-683-4678.
- Quantity in commerce
- 200
- Distribution
- Worldwide Distribution -- United States, India, Japan and Slovenia.
- Product codes and lots
- M00566530, Lot Numbers: 11941297 and 11996705; and M00566531, Lot Numbers: 11941297 and 11996705.