Z-1619-2009

Triathlon Revision Instruments, 3-6 TCG Upper Tray; Non Sterile; Howmedica Osteonics Corp., 325 Corporate...

Howmedica Osteonics Corp / initiated 2009-05-15 / Terminated

Stryker Howmedica Osteonics Corp. began a recall of Triathlon Revision Instruments, 3-6 TCG Upper Tray; Non Sterile; Howmedica Osteonics Corp., 325 Corporate Drive, Mahwah, NJ 07430. For total knee arthroplasty... on . The reason FDA records is "The screen print for the Valgus Adapter catalog numbers on the Triathlon Trial Cutting Guide Instrument Trays is incorrect". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-1619-2009
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Howmedica Osteonics Corp
Product
Triathlon Revision Instruments, 3-6 TCG Upper Tray; Non Sterile; Howmedica Osteonics Corp., 325 Corporate Drive, Mahwah, NJ 07430. For total knee arthroplasty procedures involving Triathlon Total Stabilizer components. The trays sterilize and transport instruments needed for such procedures.
Product code
JWH Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer
Reason for recall
The screen print for the Valgus Adapter catalog numbers on the Triathlon Trial Cutting Guide Instrument Trays is incorrect.
FDA root cause
Process change control
Action
Stryker Orthopaedics issued an "Urgent Product Correction" notice dated May 15, 2009 via Federal Express addressed to Branch/Agency Manager/Quality Contact informing them of the affected device. The letter outlined instructions for correction and requested consignees to complete and return a Product Correction Acknowledgment Form by fax to 1-201-831-6069. Questions can be directed to Colleen O'Meara, Manager, Divisional Regulatory Reporting at 201-831-5970.
Quantity in commerce
26
Distribution
Worldwide Distribution -- US and Canada.
Product codes and lots
Catalog Number: 6543-8-014.