Phoenix x/ray nanotom series. Phoenix x|ray nanotom series cabinet x-ray systems are used in non-destructive...
GE Inspection Technologies, LP began a recall of Phoenix x/ray nanotom series. Phoenix x|ray nanotom series cabinet x-ray systems are used in non-destructive testing, failure analysis, process control, and... on . The reason FDA records is "Some of the systems do not have a safety interlock on the service access panel. Although the systems have two independent means for indicating when x-rays are being generated, some systems may have one indicator that is...". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.
Recall record
- Recall number
- Z-1828-2011
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- GE Inspection Technologies, LP
- Product
- Phoenix x/ray nanotom series. Phoenix x|ray nanotom series cabinet x-ray systems are used in non-destructive testing, failure analysis, process control, and metrology.
- Product code
- RCE Cabinet X-Ray, Industrial
- Reason for recall
- Some of the systems do not have a safety interlock on the service access panel. Although the systems have two independent means for indicating when x-rays are being generated, some systems may have one indicator that is not discernible. Also, some systems lack proper placement of a caution label and others do not have this label affixed.
- Root cause tag
- labeling and UDI
- FDA root cause
- Radiation Control for Health and Safety Act
- Action
- The firm has completed all of the following necessary actions and no further actions have been indicated: A Field Service Memo was issued to customers. It described the affected product and described the issue. GE Sensing and Inspection provided and installed retrofit kits, and the systems in question were tested by qualified Field Service Engineers at no charge to customers. GE asked customers to post the notification letter on the affected equipment until retrofit was completed. Questions were to be directed to the GE Service Center at 800-766-6062.
- Quantity in commerce
- 118 units total
- Distribution
- Nationwide Distribution
- Product codes and lots
- All affected units.