Z-1931-2009
iPulse Circulatory Support System Catalog number: 0036-0010 The iPulse System Console is a multi-purpose...
Abiomed, Inc. / initiated 2009-04-23 / Terminated / root cause: design
Abiomed, Inc. began a recall of iPulse Circulatory Support System Catalog number: 0036-0010 The iPulse System Console is a multi-purpose mechanical circulatory support system used to support... on . The reason FDA records is "iPulse Console SC1035 displayed a "Low Pressure / Low Flow" alarm and stopped pumping during patient transport in battery operation". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-1931-2009
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Abiomed, Inc.
- Product
- iPulse Circulatory Support System Catalog number: 0036-0010 The iPulse System Console is a multi-purpose mechanical circulatory support system used to support patients suffering from acute cardiovascular failure. The iPulse console can be used to drive exclusively one of two classes of therapeutic devices: either pneumatically actuated Ventricular Assist Devices (VADs) for total circulatory support, or (i.e., non-simultaneously) an intra-aortic balloon (IAB) catheter for counter-pulsation therapy.
- Product code
- DSQ Ventricular (Assist) Bypass
- Reason for recall
- iPulse Console SC1035 displayed a "Low Pressure / Low Flow" alarm and stopped pumping during patient transport in battery operation.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- An "Urgent Voluntary Medical Device Field Correction Notice" was issued April 8, 2009 to affected customers. The customers were notified that the iPulse Console SC1032 displayed a "Low Pressure/ Low Flow" alarm and stopped pumping during patient transport in battery operation. Instructions were provided to users in case they encounter this failure mode. ABIOMED Field Service Engineers will contact customers to schedule a time to visit site and correct affected consoles. Please contact ABIOMED's Field Service at 1.800.554.8666, if you have any questions.
- Quantity in commerce
- 79
- Distribution
- Domestic: MA, MI, SC, NY, NJ, OH, CA, PA, AR, MI, MD, CT, IL, WI, TX, NJ, TN, IN, MI, DE, FL, IA, WA, KY, and WV Foreign: Germany, Greece, Italy, Latvia, Spain, and Turkey
- Product codes and lots
- Serial numbers: SC1000, SC1001, SC1003 - SC1010, SC1014 - SC1023, SC1025 - SC1070, SC1072 - SC1081, SC1083, SC1084, and SC1089