Multi-Drug iCup Drug Screen Test Cup. Part Number DUE-1127-022-08A. Subsequent Product Codes: DIO, DIS, DJC...
Biosite Inc began a recall of Multi-Drug iCup Drug Screen Test Cup. Part Number DUE-1127-022-08A. Subsequent Product Codes: DIO, DIS, DJC, DJG, DJR, JXM, JXN, LCM, LDJ, LFG on . The reason FDA records is "The cup associated with these lots of product may have a pin-sized hole in the floor of the cup. As a result, there is a potential for affected lots of cups to leak urine after an extended period of time". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0009-2010
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Alere San Diego, Inc.
- Product
- Multi-Drug iCup Drug Screen Test Cup. Part Number DUE-1127-022-08A. Subsequent Product Codes: DIO, DIS, DJC, DJG, DJR, JXM, JXN, LCM, LDJ, LFG
- Product code
- DKZ Enzyme Immunoassay, Amphetamine
- Reason for recall
- The cup associated with these lots of product may have a pin-sized hole in the floor of the cup. As a result, there is a potential for affected lots of cups to leak urine after an extended period of time.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- Biosite, Inc. issued a "Notice of Correction" letter dated May 1, 2009 to consignees describing the problem and instructed users of proper use in order to avoid any cross contamination of samples until all affected product is completely used. For further information, contact Biosite, Inc. using the following information: Biosite Incorporated, an Inverness Medical Company 9975 Summers Ridge Road San Diego, CA 92121 Phone: 1-877-441-7440, option 2 Fax: 1-858-695-7100 Email: [email protected]
- Quantity in commerce
- 113,575 cups distributed in total
- Distribution
- Nationwide Distribution
- Product codes and lots
- Lot Numbers: DOA8090784, DOA8090785, DOA8090786, DOA8090787 and DOA8110567.