Multi-Drug 8 Panel Integrated iCup Drug Screen Test Cup. Part Number DUE-187-071-08A. Subsequent Product...

Alere San Diego, Inc. / initiated 2009-04-27 / Terminated / root cause: manufacturing process

Biosite Inc began a recall of Multi-Drug 8 Panel Integrated iCup Drug Screen Test Cup. Part Number DUE-187-071-08A. Subsequent Product Codes: DIO, DIS, DJC, DJG, DJR, JXM, JXN, LCM, LDJ, LFG on . The reason FDA records is "The cup associated with these lots of product may have a pin-sized hole in the floor of the cup. As a result, there is a potential for affected lots of cups to leak urine after an extended period of time". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0010-2010
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Alere San Diego, Inc.
Product
Multi-Drug 8 Panel Integrated iCup Drug Screen Test Cup. Part Number DUE-187-071-08A. Subsequent Product Codes: DIO, DIS, DJC, DJG, DJR, JXM, JXN, LCM, LDJ, LFG
Product code
DKZ Enzyme Immunoassay, Amphetamine
Reason for recall
The cup associated with these lots of product may have a pin-sized hole in the floor of the cup. As a result, there is a potential for affected lots of cups to leak urine after an extended period of time.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
Biosite, Inc. issued a "Notice of Correction" letter dated May 1, 2009 to consignees describing the problem and instructed users of proper use in order to avoid any cross contamination of samples until all affected product is completely used. For further information, contact Biosite, Inc. using the following information: Biosite Incorporated, an Inverness Medical Company 9975 Summers Ridge Road San Diego, CA 92121 Phone: 1-877-441-7440, option 2 Fax: 1-858-695-7100 Email: [email protected]
Quantity in commerce
113,575 cups distributed in total
Distribution
Nationwide Distribution
Product codes and lots
Lot Numbers: DOA8060786, DOA8060787 and DOA8090822.