Z-1946-2009

ARJO Carendo Multi-Purpose Hygiene Chair; Model BIB2003-01, a non-AC powered patient lift; Manufactured by...

Arjohuntleigh AB / initiated 2009-06-11 / Terminated / root cause: design

ArjoHuntleigh began a recall of ARJO Carendo Multi-Purpose Hygiene Chair; Model BIB2003-01, a non-AC powered patient lift; Manufactured by Arjo Hospital Equipment AB, Eslov, Sweden. The... on . The reason FDA records is "Male patient genitals may be pinched between the seat and the frame of the patient lift chair when the Care Raiser seat is returned to its lowest position". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-1946-2009
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Arjohuntleigh AB
Product
ARJO Carendo Multi-Purpose Hygiene Chair; Model BIB2003-01, a non-AC powered patient lift; Manufactured by Arjo Hospital Equipment AB, Eslov, Sweden. The equipment is intended for hygiene care, showering, and toileting of hospital or nursing home residents under the supervision of trained nursing staff and in accordance with the instructions outlined in the Operating and Product Care Instructions.
Product code
FSA Lift, Patient, Non-Ac-Powered
Reason for recall
Male patient genitals may be pinched between the seat and the frame of the patient lift chair when the Care Raiser seat is returned to its lowest position.
Root cause tag
design
FDA root cause
Device Design
Action
ArjoHuntleigh sent Urgent Device Field Correction letters dated June 11, 2009, along with the Urgent Field Safety Notice ID 090526AHE dated May 26, 2009, via UPS Ground delivery. Customers were informed of the affected device and were requested to complete the response form with new contact information and fax it back to ArjoHuntleigh at 630-567-5012 or 630-925-7969. For further questions, contact ArjoHuntleigh at 1-800-323-1245.
Quantity in commerce
1,987 hygiene chairs
Distribution
Nationwide Distribution
Product codes and lots
Model Number: BIB2003-01, all Serial Numbers below 0905000031.