Z-1967-2012Class II

Edwards Lifesciences Rigid Suction Wand, sterile. Model numbers S033 and SPC2081

Edwards Lifesciences LLC / initiated 2012-06-06 / Terminated / root cause: manufacturing process

Edwards Lifesciences, LLC began a recall of Edwards Lifesciences Rigid Suction Wand, sterile. Model numbers S033 and SPC2081. Product Usage: These suction catheters are typically used to remove fluid on... on . FDA classified it as Class II and gives the reason as "Edwards is recalling Suction Wand models SPC2081 and S033 because the plastic tip attached to the metal suction wand may become detached during routine use". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1967-2012
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Edwards Lifesciences LLC
Product
Edwards Lifesciences Rigid Suction Wand, sterile. Model numbers S033 and SPC2081. Product Usage: These suction catheters are typically used to remove fluid on the external surface of the heart during cardiac surgery and therefore there is little risk of intravascular involvement.
Product code
DWF Catheter, Cannula And Tubing, Vascular, Cardiopulmonary Bypass
Reason for recall
Edwards is recalling Suction Wand models SPC2081 and S033 because the plastic tip attached to the metal suction wand may become detached during routine use.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
Edwards Lifesciences sent an Urgent Field Safety Notice Product Recall letter dated June 2012 to all affected customers. The letter identified the affected product, description of the issue and actions to be taken. Customers were instructed to examine their inventory for the affected product, quarantine and return any remaining in stock and complete the enclosed acknowledgment form with fax instructions to assure that the notification was received. For question call Edwards Customer Service at (800) 424-3278 from 8:00AM-4:30PM Pacific Time.
Quantity in commerce
1524 units
Distribution
Worldwide Distribution - US (nationwide) including the states of CT, TX, CA, AZ, MO, and MD and the countries of Europe and Japan.
Product codes and lots
Lot numbers: 58749813, 58802303, 58836847, 58877405, 58909346, 59011910, 59035360, 59085935, 59118391, 59241531, 58705682, 58711964, 58722234, 58771175, 58837337, 58855247, 58904182, 58927518, 58936037, 58936014, 58936027, 59009635, 59030253, 59056470, 59103213, 59124020, 59199172, 59218725, 59235997.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1967-2012
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1967-2012Class IITerminatedVoluntary: Firm initiated