Z-1653-2009

Arrow Pediatric Two-Lumen Central Venous Catheterization Set with Blue FlexTip Catheter, 5 Fr., 2 Lumen, 13cm...

Arrow International LLC (subsidiary of Teleflex Incorporated) / initiated 2009-06-17 / Terminated / root cause: manufacturing process

Arrow International Inc began a recall of Arrow Pediatric Two-Lumen Central Venous Catheterization Set with Blue FlexTip Catheter, 5 Fr., 2 Lumen, 13cm catheter length, Catalog number: CS-14502 on . The reason FDA records is "The spring wire guide is too large in diameter to pass through the needle, dilator and catheter". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1653-2009
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Arrow International LLC (subsidiary of Teleflex Incorporated)
Product
Arrow Pediatric Two-Lumen Central Venous Catheterization Set with Blue FlexTip Catheter, 5 Fr., 2 Lumen, 13cm catheter length, Catalog number: CS-14502.
Product code
DQY Catheter, Percutaneous
Reason for recall
The spring wire guide is too large in diameter to pass through the needle, dilator and catheter.
Root cause tag
manufacturing process
FDA root cause
Process design
Action
Arrow notified consignees by letter on/about 06/17/2009. They were instructed to cease use and distribution and quarantine all affected product. They were to contact Arrow's Customer Service Department for a Return Authorization Number. Enclosed was a Recall Acknowledgement & Stock Status Form to be completed and faxed to Arrow International to verify receipt of notification and acknowledge the amount of return.
Quantity in commerce
3019 units
Distribution
Worldwide distribution, including USA, Argentina, Brazil, India, Taiwan and Thailand.
Product codes and lots
Catalog number: CS-14502, Lot numbers: RF7019488 and RF7029944.