Hitachi CXR4 Computed Tomography (CT) Scanners, Software Version 1.51 or less
Hitachi Medical Systems America Inc began a recall of Hitachi CXR4 Computed Tomography (CT) Scanners, Software Version 1.51 or less on . The reason FDA records is "Incorrect scale on image: A software error can occur if two (2) different Field-of-View (FOV) settings are registered in a protocol using Multi-Reconstruction. This could cause the image generated to be displayed with...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-1651-2009
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- FUJIFILM Healthcare Americas Corporation
- Product
- Hitachi CXR4 Computed Tomography (CT) Scanners, Software Version 1.51 or less.
- Product code
- JAK System, X-Ray, Tomography, Computed
- Reason for recall
- Incorrect scale on image: A software error can occur if two (2) different Field-of-View (FOV) settings are registered in a protocol using Multi-Reconstruction. This could cause the image generated to be displayed with the wrong scale factor rendering anatomical measurements incorrect.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- HMSA issued an URGENT - DEVICE CORRECTION notification to their customers on 4/14/2009. The notification describes the potential software error and the condition under which these image distortions can occur. The notification goes on to provide the customer with a work-around procedure which they can use to avoid encountering this issue with their current software configuration(s). The letter states that the firm will be installing updated software on all affected systems beginning in June, 2009, and that a Hitachi Service Rep. will be contacting each customer and making arrangements to come out and install a software upgrade on the affected system(s) at no cost to the customer. Lastly, the letter provides a telephone # to the firm's Applications Helpline: 1-877-228-2777, which the customer can use to obtain additional information, if desired.
- Quantity in commerce
- 47 units
- Distribution
- Nationwide distribution, including Puerto Rico (PR).
- Product codes and lots
- The recalled CXR4 system units are identified with the following Code Numbers: CXR46201 through CXR4622 and CXR4624 through CXR46244.