Z-1711-2009

Laborie Medical Technologies Urodynamic Catheters, Twin Lumen 6 Fr , 40cm For Cystometry. Catalog Number...

Abbott Medical / initiated 2009-04-09 / Terminated / root cause: packaging

Laborie Medical Technologies Cor began a recall of Laborie Medical Technologies Urodynamic Catheters, Twin Lumen 6 Fr , 40cm For Cystometry. Catalog Number: CAT-206 on . The reason FDA records is "Laborie Medical Technologies reports that there was a mix up in the labeling and packaging of their Urodynamic Catheters. There were 8Fr catheters mixed inside the 6Fr catheter packages". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.

Recall record

Recall number
Z-1711-2009
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Abbott Medical
Product
Laborie Medical Technologies Urodynamic Catheters, Twin Lumen 6 Fr , 40cm For Cystometry. Catalog Number: CAT-206.
Product code
EXQ Cystometer, Electrical Recording
Reason for recall
Laborie Medical Technologies reports that there was a mix up in the labeling and packaging of their Urodynamic Catheters. There were 8Fr catheters mixed inside the 6Fr catheter packages.
Root cause tag
packaging
FDA root cause
Packaging process control
Action
Laborie Medical Technologies notified customers by telephone/email beginning April 10, 2009 advising them of the mislabelled product. For further questions, contact Laborie Medical Technologies at 1-800-522-6743.
Quantity in commerce
28 boxes (10's)
Distribution
Nationwide Distribution -- CA, GA, IL, KS, MD, PA, OH, OK, UT and WA.
Product codes and lots
Lot Number: 82484.