Z-0421-2010
Kuschall K Junior Manual wheelchair
Invacare GmbH / initiated 2009-06-15 / Terminated / root cause: design
Invacare Corporation began a recall of Kuschall K Junior Manual wheelchair on . The reason FDA records is "The handle has been breaking at a calculated weight of 65 kg which may have serious consequences (broken bones) depending on the circumstances of the break (e.g., user on ramps, hills, etc.)". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-0421-2010
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Invacare GmbH
- Product
- Kuschall K Junior Manual wheelchair
- Product code
- IOR Wheelchair, Mechanical
- Reason for recall
- The handle has been breaking at a calculated weight of 65 kg which may have serious consequences (broken bones) depending on the circumstances of the break (e.g., user on ramps, hills, etc.).
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- The firm contacted their customers by telephone and then sent a recall notification letter to each customer on 6/15/2009. Questions are directed to email: [email protected] or phone: 800-333-6900.
- Quantity in commerce
- 20 Wheelchairs
- Distribution
- Product was distributed to the following states: CA, CO, MN, NJ, OR, PA, TX & VA and one to Canada.
- Product codes and lots
- Model Numbers: 0806281234, 0806893858, 0806135534, 0806161578, 0806237627, 0806268153, 0806268154, 0806309590, 0806315524, 0806351557, 0806568768, 0806591793. 0806670117, 0806724855, 0806731287, 0806773921, 0807007811, 0807019897, 0807062866, 0807110193