Z-2005-2009
Roche ACCU-CHEK Performa blood glucose test strips, Roche Diagnostics, Indianapolis, IN. Test strips for...
Roche Diagnostics GmbH / initiated 2009-06-10 / Terminated / root cause: manufacturing process
Roche Diagnostics Corp. began a recall of Roche ACCU-CHEK Performa blood glucose test strips, Roche Diagnostics, Indianapolis, IN. Test strips for quantitative blood glucose measurements in fresh... on . The reason FDA records is "There may be a significant negative bias when testing neonate samples". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-2005-2009
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Roche Diagnostics GmbH
- Product
- Roche ACCU-CHEK Performa blood glucose test strips, Roche Diagnostics, Indianapolis, IN. Test strips for quantitative blood glucose measurements in fresh capillary, arterial, or neonatal blood, as well as in heparin (lithium or sodium) or EDTA anticoagulated venous blood.
- Product code
- NBW System, Test, Blood Glucose, Over The Counter
- Reason for recall
- There may be a significant negative bias when testing neonate samples.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- Roche Diagnostics notified consignees in the affected countries of the problem by "Urgent Field Safety Notice" dated June 10, 2009 and instructed to cease using the Accu-Chek Performa and Accu-Chek Inform II systems on neonates, except for systems using test strips with advanced chemistry. For questions and additional information, contact your local Roche Diagnostics sales affiliate or sales representative.
- Quantity in commerce
- 1,518 of total products
- Distribution
- International Distribution -- Argentina, Australia, Brazil, Chili, China, Colombia, Ecuador, France, Germany, Great Britain, Hong Kong, India, Korea, Malaysia, Mexico, New Zealand, Pakistan, Panama, Peru, Philippines, Singapore, South Africa, Taiwan, Thailand, Uruguay and Venezuela.
- Product codes and lots
- All Lots with REF Numbers: 4861680020, 4861680023, 4861680031, 4861680039, 4861680053, 4861680136, 4862414003, 4862414016, 4862414020, 4862414023, 4862414031, 4862414039, 4862414053, 4862414136, 4862414170, 4948858020, 4948858039, 4948874031 and 5235243001.