Hospira Phoenix Infusion System with MedNet Software, Symbiq One-Channel Infusion System, Software versions...
Hospira Inc began a recall of Hospira Phoenix Infusion System with MedNet Software, Symbiq One-Channel Infusion System, Software versions v2.1 & v3.0, Manufactured by Hospira Inc., Morgan... on . The reason FDA records is "Potential delay/underinfusion of critical therapy-- Devices experience increased frequency of alarms that result in temporary pump stoppage". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-1815-2009
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Abbott Medical
- Product
- Hospira Phoenix Infusion System with MedNet Software, Symbiq One-Channel Infusion System, Software versions v2.1 & v3.0, Manufactured by Hospira Inc., Morgan Hill, CA.
- Product code
- frn
- Reason for recall
- Potential delay/underinfusion of critical therapy-- Devices experience increased frequency of alarms that result in temporary pump stoppage.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- Hospira, Inc. issued an "Urgent: Device Field Correction" dated March 31, 2009 via Federal Express to Healthcare Professionals and Valued Hospira Customers. Consignees were informed of the affected devices and will be contacted by a Hospira representative to arrange necessary device upgrades. For further questions, contact Hospira Global Product Safety and Complaints at 1-800-441-4100.
- Distribution
- Product was distributed to 59 consignees to all 50 States.
- Product codes and lots
- List number 16026-04-83/84 and 87/88, software versions 2.1 and 3.0 only.