Stryker PainPump2, 400 mL PainPump with luer lock tubing set, 5.0 inch ExFen and standard catheter, sterile...
Stryker Instruments Div. of Stryker Corporation began a recall of Stryker PainPump2, 400 mL PainPump with luer lock tubing set, 5.0 inch ExFen and standard catheter, sterile, Stryker Instruments, Kalamazoo, MI; REF 540-155 on . The reason FDA records is "The catheter may break when removed from the body, or the catheter may have kinks or be occluded such that the patient will not receive the intended quantity of medication". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1780-2009
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Stryker Corp.
- Product
- Stryker PainPump2, 400 mL PainPump with luer lock tubing set, 5.0 inch ExFen and standard catheter, sterile, Stryker Instruments, Kalamazoo, MI; REF 540-155.
- Product code
- FRN Pump, Infusion
- Reason for recall
- The catheter may break when removed from the body, or the catheter may have kinks or be occluded such that the patient will not receive the intended quantity of medication.
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- Stryker Instruments issued a recall letter dated June 16, 2009 and instructed Consignees to cease using and destroy the product and to return a business reply form for credit via fax to Stryker Instruments Regulatory Department, 1-866-521-2762. For further questions, contact Stryker Instruments at 1-800-800-4236 extension 3808.
- Quantity in commerce
- 2634
- Distribution
- Worldwide Distribution -- Including the United States, Argentina, Australia, Brazil, Canada, England, France, Germany, Hong Kong, Italy, Japan, Latin America, Malaysia, Mexico, Netherlands, New Zealand, Singapore, South Africa, Spain, Sweden, Switzerland and Taiwan.
- Product codes and lots
- Lot Numbers: 07054012 through 09077012 and 6033374026 through 7017529998.