Z-1895-2009

Boston Scientific Clipping Device, Resolution Clip, working length 155 cm, working channel 2.8 mm...

Boston Scientific Corporation / initiated 2009-06-17 / Terminated / root cause: manufacturing process

Boston Scientific Corporation began a recall of Boston Scientific Clipping Device, Resolution Clip, working length 155 cm, working channel 2.8 mm, Manufactured for Boston Scientific Corporation, Natick, MA... on . The reason FDA records is "The clip may not deploy or, if deployed, the clip may not release from the tissue". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1895-2009
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Boston Scientific Corporation
Product
Boston Scientific Clipping Device, Resolution Clip, working length 155 cm, working channel 2.8 mm, Manufactured for Boston Scientific Corporation, Natick, MA; Product codes M00522600 (1 clip) and M00522602 (20 clips). The Resolution Clip Device is a pre-loaded single-use clip on a flexible delivery system. It is placed into the gastrointestinal tract for the purpose of endoscopic marking, hemostasis, anchoring jejunal tubes and closure of lumenal perforations.
Product code
FHN Ligator, Hemorrhoidal
Reason for recall
The clip may not deploy or, if deployed, the clip may not release from the tissue.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
Consignees were notified by an "Urgent Recall Letter " dated June 17, 2009 and instructed to cease using the product because the calibration of equipment used to manufacture these products was not in control. This may potentially result in an inability to deploy or release the clip from the catheter. Consignees were instructed to return a business reply form for return authorization. Please contact Cindy Finney, Field Action Team Lead at 1-508-683-4678 for assistance.
Quantity in commerce
217
Distribution
Nationwide, France, Germany, Great Britain, Italy, Netherlands, Spain and Sweden.
Product codes and lots
Product code M00522600, lot 0ML9042402 and Product code M00522602, lot 0ML9042701.