Z-1896-2009
Boston Scientific Clipping Device, Resolution Clip, working length 235 cm, working channel 2.8 mm...
Boston Scientific Corporation / initiated 2009-06-17 / Terminated / root cause: manufacturing process
Boston Scientific Corporation began a recall of Boston Scientific Clipping Device, Resolution Clip, working length 235 cm, working channel 2.8 mm, Manufactured for Boston Scientific Corporation, Natick, MA... on . The reason FDA records is "The clip may not deploy or, if deployed, the clip may not release from the tissue". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1896-2009
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Boston Scientific Corporation
- Product
- Boston Scientific Clipping Device, Resolution Clip, working length 235 cm, working channel 2.8 mm, Manufactured for Boston Scientific Corporation, Natick, MA; Product codes M00522610 (1 clip) and M00522612 (20 clips). The Resolution Clip Device is a pre-loaded single-use clip on a flexible delivery system. It is placed into the gastrointestinal tract for the purpose of endoscopic marking, hemostasis, anchoring jejunal tubes and closure of lumenal perforations.
- Product code
- FHN Ligator, Hemorrhoidal
- Reason for recall
- The clip may not deploy or, if deployed, the clip may not release from the tissue.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- Consignees were notified by an "Urgent Recall Letter " dated June 17, 2009 and instructed to cease using the product because the calibration of equipment used to manufacture these products was not in control. This may potentially result in an inability to deploy or release the clip from the catheter. Consignees were instructed to return a business reply form for return authorization. Please contact Cindy Finney, Field Action Team Lead at 1-508-683-4678 for assistance.
- Quantity in commerce
- 4269
- Distribution
- Nationwide, France, Germany, Great Britain, Italy, Netherlands, Spain and Sweden.
- Product codes and lots
- Product code M00522610, lots 0ML9042004, 0ML9042101, 0ML9042102, 0ML9042702, 0ML9050401 and 0ML9051201. Product code M00522612, lots 0ML9042005, 0ML9042006, 0ML9042204, 0ML9042705, 0ML9042706, 0ML9050402, 0ML9050403, 0ML9051205 and 0ML9051206.