Z-1811-2009
BD 60ml Syringe Luer-Lok tip, Sterile, BD, Franklin Lakes, NJ 07417, Made in USA
Becton Dickinson & Co / initiated 2009-07-07 / Terminated / root cause: packaging
Becton Dickinson & Company began a recall of BD 60ml Syringe Luer-Lok tip, Sterile, BD, Franklin Lakes, NJ 07417, Made in USA on . The reason FDA records is "Insufficient Seal: Becton Dickinson received complaints of insufficient/open seals on one lot of 60 ml Luer-Lok Syringes". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.
Recall record
- Recall number
- Z-1811-2009
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Becton Dickinson & Co
- Product
- BD 60ml Syringe Luer-Lok tip, Sterile, BD, Franklin Lakes, NJ 07417, Made in USA.
- Product code
- FMF Syringe, Piston
- Reason for recall
- Insufficient Seal: Becton Dickinson received complaints of insufficient/open seals on one lot of 60 ml Luer-Lok Syringes.
- Root cause tag
- packaging
- FDA root cause
- Packaging process control
- Action
- Becton Dickinson issued an "Urgent: Product Recall" notice via UPS tracking dated July 7, 2009 to all distributors informing them of the affected product. The firm asked all distributors to identify end customers in an attempt to notify all customers. The firms asks that all affected product be returned. Further questions should be directed to BD Customer Service representatives at 1-888-237-2762.
- Quantity in commerce
- 133,600 units
- Distribution
- Nationwide Distribution -- Including states of IL, CA, VA, PA, and FL.
- Product codes and lots
- Product Number: 309653; Lot Number: 819101A.