Z-1811-2009

BD 60ml Syringe Luer-Lok tip, Sterile, BD, Franklin Lakes, NJ 07417, Made in USA

Becton Dickinson & Co / initiated 2009-07-07 / Terminated / root cause: packaging

Becton Dickinson & Company began a recall of BD 60ml Syringe Luer-Lok tip, Sterile, BD, Franklin Lakes, NJ 07417, Made in USA on . The reason FDA records is "Insufficient Seal: Becton Dickinson received complaints of insufficient/open seals on one lot of 60 ml Luer-Lok Syringes". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.

Recall record

Recall number
Z-1811-2009
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Becton Dickinson & Co
Product
BD 60ml Syringe Luer-Lok tip, Sterile, BD, Franklin Lakes, NJ 07417, Made in USA.
Product code
FMF Syringe, Piston
Reason for recall
Insufficient Seal: Becton Dickinson received complaints of insufficient/open seals on one lot of 60 ml Luer-Lok Syringes.
Root cause tag
packaging
FDA root cause
Packaging process control
Action
Becton Dickinson issued an "Urgent: Product Recall" notice via UPS tracking dated July 7, 2009 to all distributors informing them of the affected product. The firm asked all distributors to identify end customers in an attempt to notify all customers. The firms asks that all affected product be returned. Further questions should be directed to BD Customer Service representatives at 1-888-237-2762.
Quantity in commerce
133,600 units
Distribution
Nationwide Distribution -- Including states of IL, CA, VA, PA, and FL.
Product codes and lots
Product Number: 309653; Lot Number: 819101A.