Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1870-2009

GE Healthcare, Uni-Circuit Kit, Adult, Disposable, 1.8 m/72 in, REF M1089028, M1087972, M1087975, M1126445...

GE Healthcare / initiated 2009-06-05 / Terminated / root cause: manufacturing process

GE Medical Systems, LLC began a recall of GE Healthcare, Uni-Circuit Kit, Adult, Disposable, 1.8 m/72 in, REF M1089028, M1087972, M1087975, M1126445, M1127508, M1127509, M1129780, M1129781, M1130230... on . The reason FDA records is "Breathing circuit inner tube may separate from the rigid connectors under certain conditions. A separation of the inner tube from the connector may prevent adequate gas flow (O2, N2O and anesthetic agents) to the...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1870-2009
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
GE Healthcare
Product
GE Healthcare, Uni-Circuit Kit, Adult, Disposable, 1.8 m/72 in, REF M1089028, M1087972, M1087975, M1126445, M1127508, M1127509, M1129780, M1129781, M1130230, M1137805, M1137199, M1145542, M1170604, M1151662, M1152224, M1158990, M1100486, M1122080, M1126444, M1163003, M1163255 M1132533, M1133966, M1170051, M1135144, M1139485, M1118627, M1118628, M1145406, M1147298, M1137811, M1152651, M1160275. pkg of 20 pcs, Latex Free. Breathing circuit for use with anesthesia machine or ventilator.
Product code
CAI Circuit, Breathing (W Connector, Adaptor, Y Piece)
Reason for recall
Breathing circuit inner tube may separate from the rigid connectors under certain conditions. A separation of the inner tube from the connector may prevent adequate gas flow (O2, N2O and anesthetic agents) to the patient. A separation can also cause CO2 re-breathing. This could lead to hypoxia and/or hypercarbia and inadequate anesthesia delivery.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
GE Healthcare sent consignees 6/5/09 an "Urgent Medical Device Correction" letter dated June 5, 2009. The letter was addressed to Health Care Administrator/Risk Manager; Chief of Anesthesia; Director of Clinical Engineering; Risk Manager. The letter described the Safety Issue, Affected Product Details, Safety Instructions (included "Recall Confirmation" form fax to 973-790-4150), Product Correction and Contact Information.
Quantity in commerce
85097
Distribution
Worldwide distribution: USA, AUSTRALIA, DENMARK, FRANCE, ESTONIA, GERMANY, ITALY, JORDAN, NETHERLANDS, NORWAY, SWEDEN, and UNITED ARAB EMIRATES.
Product codes and lots
This correction pertains to Uni-Circuits from lot# 084812 and later, up to and including lot# 092210