Z-2027-2009
Smith & Nephew Orthopaedics PIP (proximal Interphalangeal) FIXATION HINGE KIT, REF 101638, for external...
Smith & Nephew, Inc. / initiated 2009-01-23 / Terminated / root cause: manufacturing process
Smith & Nephew Inc began a recall of Smith & Nephew Orthopaedics PIP (proximal Interphalangeal) FIXATION HINGE KIT, REF 101638, for external fixation of finger joint on . The reason FDA records is "Hinge failure: The product is cracking or breaking during use". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-2027-2009
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Smith & Nephew, Inc.
- Product
- Smith & Nephew Orthopaedics PIP (proximal Interphalangeal) FIXATION HINGE KIT, REF 101638, for external fixation of finger joint.
- Product code
- JDW Pin, Fixation, Threaded
- Reason for recall
- Hinge failure: The product is cracking or breaking during use.
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- The Smith & Nephew Sales Representatives were notified of the problem and the recall via certified letter and e-mail on 01/23/2009. A second notice was sent to the hospitals on 06/06/2009. All affected Smith & Nephew International Distributors were notified via email on 01/23/2009. They were instructed to immediately notify their accounts concerning the recall. Smith & Nephew c/o Return Goods, 4287 Delp St, Memphis, TN 38118
- Quantity in commerce
- 7,089 units
- Distribution
- Worldwide Distribution: USA, Turkey, Denmark, Germany, Malaysia, Hong Kong, Korea, Switzerland, Japan, Singapore, Canada, New Zealand, Australia, and Italy.
- Product codes and lots
- All Lots
FIRST ARTICLE / WEEKLY