Sterrad CycleSure Biological Indicator, Product Code 14324. The Sterrad CycleSure Biological Indicator is...
Advanced Sterilization Products began a recall of Sterrad CycleSure Biological Indicator, Product Code 14324. The Sterrad CycleSure Biological Indicator is intended to be used as a standard method for frequent... on . The reason FDA records is "Biological Indicator indicating a positive result does not maintain yellow color for the entire incubation period and some indicators experience media evaporation prior to the end of the labeled maximum incubation time". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-0066-2010
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Advanced Sterilization Products
- Product
- Sterrad CycleSure Biological Indicator, Product Code 14324. The Sterrad CycleSure Biological Indicator is intended to be used as a standard method for frequent monitoring of the Sterrad Sterilizer cycles.
- Product code
- FRC Indicator, Biological Sterilization Process
- Reason for recall
- Biological Indicator indicating a positive result does not maintain yellow color for the entire incubation period and some indicators experience media evaporation prior to the end of the labeled maximum incubation time.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- An Important Product Labeling Correction - New Instructions for Use dated, June 26, 2009, has been proposed to be sent by the firm for Cyclosure Biological Indicators to consignees. The revised instructions reduce the maximum read time for the indicators to 3 days (72 hours). The letter also states why ASP changed the Instructions. Questions are to be directed to ASP Customer Care Center.
- Distribution
- Worldwide Distribution.
- Product codes and lots
- All Lots.