Z-0169-2010

Byrd Screw, 2.0X6MM SD, Model Number: 01-0264 Product labeling reads in part as follows: REF: 01-0264...

Biomet Microfixation / initiated 2009-02-19 / Terminated / root cause: manufacturing process

Biomet Microfixation, Inc. began a recall of Byrd Screw, 2.0X6MM SD, Model Number: 01-0264 Product labeling reads in part as follows: REF: 01-0264, QTY,UOM 00001, BYRD SCREW, 2 OX6MM SELF-DRILLING, 12CM... on . The reason FDA records is "Byrd Screw, Self Drilling, may break during usage. BIOMET has determined that the parts were not correctly processed by the vendor , i.e. that the parts were not hardened in the final processing and that lead to weaker...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0169-2010
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Biomet Microfixation
Product
Byrd Screw, 2.0X6MM SD, Model Number: 01-0264 Product labeling reads in part as follows: REF: 01-0264, QTY,UOM 00001, BYRD SCREW, 2 OX6MM SELF-DRILLING, 12CM, (4-3/4) (L), STAINLESS STEEL, LOT 999999, (01) 00841036007157(10)999999(13)090201(30)13, 2009-02, ***NON-STERILE PRODUCT***DD 14 03 1, BIOMET MICROFIXATION, INC., 150 TRADEPORT DRIVE, JACKSONVILLE, FL 32218 USA Instrument used to pull up the broken bone, so it is flush and can be plated fixated. The Byrd screw is then removed.
Product code
LXH Orthopedic Manual Surgical Instrument
Reason for recall
Byrd Screw, Self Drilling, may break during usage. BIOMET has determined that the parts were not correctly processed by the vendor , i.e. that the parts were not hardened in the final processing and that lead to weaker than acceptable material.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
Biomet plans to send letters to customers on February 19, 2009. Customers are told to discontinue use of the item and return it to Biomet Microfixation. The firm plans to follow up with an e-mail to their customers. Questions pertaining to this issue, should be directed to Christy Cain at 1.800.874.7711 or 904.741.4400.
Quantity in commerce
22
Distribution
AZ, CA, GA, IL, IN, NM, NY, TX, VA, Puerto Rico and Canada.
Product codes and lots
Lot # 091807107