Z-1810-2009
CyberKnife Treatment Delivery System, a subsystem of the CyberKnife Robotic Radiosurgery System, manufactured...
Accuray Incorporated / initiated 2009-06-22 / Terminated / root cause: software
Accuray Inc began a recall of CyberKnife Treatment Delivery System, a subsystem of the CyberKnife Robotic Radiosurgery System, manufactured by Accuray Inc., Sunnyvale, CA. Indicated for... on . The reason FDA records is "Targeting accuracy out of specification, Error alert does not render the system down, which may result in mistreatment in the wrong area". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-1810-2009
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Accuray Incorporated
- Product
- CyberKnife Treatment Delivery System, a subsystem of the CyberKnife Robotic Radiosurgery System, manufactured by Accuray Inc., Sunnyvale, CA. Indicated for treatment for planning and image-guided stereotactic radiosurgery and precision radiotherapy of lesions, tumors and conditions anywhere in the body when radiation treatment is indicated.
- Product code
- IYE Accelerator, Linear, Medical
- Reason for recall
- Targeting accuracy out of specification, Error alert does not render the system down, which may result in mistreatment in the wrong area.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- Accuray, Inc. issued an "Urgent Device Correction" notice dated June 22, 2009. The consignees were informed of the defect including methods for detecting the error. The firm is in the process of developing a software patch to address the issue. For further information, contact Accuray, Inc. Customer Support at +1-877-668-8667 (US) or +1-408-716-4700 (outside US) or [email protected].
- Quantity in commerce
- 61 units
- Distribution
- Worldwide Distribution -- US including states of MA, FL, AL, TN, DE, CA, NJ, OK, CO, NC, IL, PA, MD, DC, MS, CN, NV, RI, SC, UT, VA, MI, NE, MO, TX and AL and countries of Canada, Russia, UK, Ukraine, Saudi Arabia, Switzerland, Turkey, Taiwan, India, China, Korea and Thailand.
- Product codes and lots
- All system Serial Numbers C128 and higher with 2005 edition Robot Controller.