Z-1917-2009
Triathlon MIS 4:1 Cutting Blocks - Size 1-8; Catalog Numbers: 6541-5-701, 6541-5-702, 6541-5-703, 6541-5-704...
Howmedica Osteonics Corp / initiated 2007-02-12 / Terminated / root cause: manufacturing process
Stryker Howmedica Osteonics Corp. began a recall of Triathlon MIS 4:1 Cutting Blocks - Size 1-8; Catalog Numbers: 6541-5-701, 6541-5-702, 6541-5-703, 6541-5-704, 6541-5-705, 6541-5-706, 6541-5-707, & 6541-5-708... on . The reason FDA records is "Triathlon Cutting Blocks did not assemble to the Triathlon Impactor/Extractor Handle". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1917-2009
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Howmedica Osteonics Corp
- Product
- Triathlon MIS 4:1 Cutting Blocks - Size 1-8; Catalog Numbers: 6541-5-701, 6541-5-702, 6541-5-703, 6541-5-704, 6541-5-705, 6541-5-706, 6541-5-707, & 6541-5-708; Triathlon MIS Instruments; Non Sterile: Howmedica Osteonics Corp. 325 Corporate Drive Mahwah, NJ 07430 The Triathlon AR 4:1 Cutting Blocks and the Triathlon MIS 4:1 Cutting Blocks are used to prepare the Femur (bone cuts) for the Triathlon femoral component.
- Product code
- LXH Orthopedic Manual Surgical Instrument
- Reason for recall
- Triathlon Cutting Blocks did not assemble to the Triathlon Impactor/Extractor Handle.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- Important Product Correction letters were sent to all branches/agencies on February 12, 2008 by Federal Express. The letter asked consignees to examine their inventory and hospital locations to identify the affected product. Customers were asked to perform a functional inspection by assembling the Triathlon Impactor/Extractor Handle to all of their MIS Cutting Blocks. If the cutting block does not assemble, customers are to list the affected lot codes on the Attached Product Accountability Form and fax a copy to (201) 831-6069. The letter states that product should not be returned until further instructions are received. Questions should be directed to (201) 831-5825.
- Quantity in commerce
- 690 - both varieties
- Distribution
- Worldwide Distribution
- Product codes and lots
- Catalog numbers: 6541-5-701, 6541-5-702, 6541-5-703, 6541-5-704, 6541-5-705, 6541-5-7-6, 6541-5-707, 6541-5-708. All lot codes.