Boston Scientific, LeVeen Needle Electrode, 15 cm, 5.0 cm, sterile; REF M001262160. For tissue coagulation
Boston Scientific Corporation / initiated 2009-07-06 / Terminated / root cause: manufacturing process
Boston Scientific Corp began a recall of Boston Scientific, LeVeen Needle Electrode, 15 cm, 5.0 cm, sterile; REF M001262160. For tissue coagulation on . The reason FDA records is "The tines may detach from the core wire during use, which may result in bleeding or soft tissue trauma requiring medical intervention". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1876-2009
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Boston Scientific Corporation
- Product
- Boston Scientific, LeVeen Needle Electrode, 15 cm, 5.0 cm, sterile; REF M001262160. For tissue coagulation.
- Product code
- GEI Electrosurgical, Cutting & Coagulation & Accessories
- Reason for recall
- The tines may detach from the core wire during use, which may result in bleeding or soft tissue trauma requiring medical intervention.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- Boston Scientific notified consignees by letter dated 7/6/09 and instructed to cease using and to return the products, and to complete and return a reply form.
- Quantity in commerce
- 2,774
- Distribution
- Nationwide.
- Product codes and lots
- Lots 8745076, 8745077, 8745078, 8745082, 8745083, 8764662, 8764663, 8840136, 8840210, 8888632, 8922046, 8922047, 8922048, 8925499, 8946944, 8952359, 8960112, 8991516, 9007921, 9023847, 9051903, 9063250, 9080980, 9103385, 9103712, 9103713, 9103714, 9135758, 9153911, 9160547, 9275203, 9288929, 9300199, 9303597, 9406392, 9447228, 9464347, 9484064, 9502208, 9516123, 9591056, 9629225, 9643197, 9659590, 9708545, 9727738, 9763201, 9776097, 9783671, 11001461, 11005784, 11034758, 11042785, 11091132, 11114016, 11128377, 11128379, 11178425, 11203868, 11230657, 11356155, 11403788, 11440970, 11460060, 11551315, 11580328, 11620899, 11711276, 11766589, 11851200, 11881851, 11900344, 11916630, 11931597, 11943359, 11947005, 12324021, 12450298, 12517661 and 12621268.