Broselow/Hinkle Pediatric Emergency System; Contents: 1 15 Ga Adjustable Length Sternal/Iliac aspiration...
Vital Signs Inc began a recall of Broselow/Hinkle Pediatric Emergency System; Contents: 1 15 Ga Adjustable Length Sternal/Iliac aspiration Needle; 1 Extension Set, Sterile. Equipment Color... on . The reason FDA records is "Not compatible: The luer connector on the Intraosseus needle provided does not fit securely on the luer connection of the IV Extension Set in the module". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-1912-2009
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Vital Signs, Inc.
- Product
- Broselow/Hinkle Pediatric Emergency System; Contents: 1 15 Ga Adjustable Length Sternal/Iliac aspiration Needle; 1 Extension Set, Sterile. Equipment Color Zone: Green, Single Patient use, INTRAOSSEUS Module, Product # 7700GIO, Vital Signs, Totowa, NJ. The Broselow/Hinkle Pediatric Emergency System is a convenience kit that contains various modules/components used for the emergency treatment of pediatric patients. The modules are color coded to accommodate patients of various sizes.
- Product code
- GAA Needle, Aspiration And Injection, Disposable
- Reason for recall
- Not compatible: The luer connector on the Intraosseus needle provided does not fit securely on the luer connection of the IV Extension Set in the module.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- Vital Signs, Inc. issued an "Urgent Medical Device Correction" letter dated June 30, 2009 to all direct and indirect accounts by first class mail. Consignees were informed of the affected device, asked to complete a Confirmation form for return and to return unused product for replacement. For further questions, contact Vital Signs, Inc. at 1-973-790-1330 extension 404.
- Quantity in commerce
- 23,429 kits
- Distribution
- Worldwide Distribution -- US, Canada, United Arab Emirate, Venezuela, Guam, Korea and United Kingdom.
- Product codes and lots
- Model Numbers: 7700BIO, 7700GIO, 7700PIO, 7700RIO, 7700WIO, 7700YIO, 7700BIO2, 7700GIO2, 7700OIO2, 7700PIO2, 7700RIO2, 7700WIO2, 7700YIO2, 7730BLU5, 7730GRN5, 7730ORG5, 7730PUR5, 7730RED5, 7730WHI5, 7730YEL5, 7730FLY, 7730MOD, 7730ALS and 7730IALS; and Lot Numbers: (2008) 245F through 366F and (2009) 001K through 149K.