American Optisurgical, Horizon Phacoemulsification System, Model: HZN-2030, for ophthalmic surgery
American Optisurgical Inc began a recall of American Optisurgical, Horizon Phacoemulsification System, Model: HZN-2030, for ophthalmic surgery on . The reason FDA records is "Device operation interrupted: A software bug causes Phaco to stop when vacuum goes above 300 mmHg while in Multi Burst mode. Use of this product at the parameters specified above could lead to a delay in surgery that...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-1808-2009
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Tenex Health
- Product
- American Optisurgical, Horizon Phacoemulsification System, Model: HZN-2030, for ophthalmic surgery.
- Product code
- HQC Unit, Phacofragmentation
- Reason for recall
- Device operation interrupted: A software bug causes Phaco to stop when vacuum goes above 300 mmHg while in Multi Burst mode. Use of this product at the parameters specified above could lead to a delay in surgery that might present a health risk.
- Root cause tag
- software
- FDA root cause
- Software change control
- Action
- American Optisurgery sent an E-mail to customers on April 2, 2009 stating the reason for recall and informed them to immediately examine their inventory and quarantine product subject to recall. In addition, if they have further distributed this product, please identify customers and identify them at once of this product recall. Their notification to customers may be enhanced by including a copy of this recall notification letter. They were then told to refer to the enclosed Recall Questionnaire for instructions on what to do with the recalled product: The recall should be carried out to the user level and return the enclosed response form as soon as possible. If there are any questions, they are advised to contact David Salzberg at (800) 576-1266.
- Quantity in commerce
- 12 units
- Distribution
- Worldwide distribution: USA (California), Brazil, Algeria, France, South Korea, India, and Peru.
- Product codes and lots
- Serial Numbers: 1114, 1116, 1118, 1120-1124, and 1126-1129.