Z-1808-2009

American Optisurgical, Horizon Phacoemulsification System, Model: HZN-2030, for ophthalmic surgery

Tenex Health / initiated 2009-03-24 / Terminated / root cause: software

American Optisurgical Inc began a recall of American Optisurgical, Horizon Phacoemulsification System, Model: HZN-2030, for ophthalmic surgery on . The reason FDA records is "Device operation interrupted: A software bug causes Phaco to stop when vacuum goes above 300 mmHg while in Multi Burst mode. Use of this product at the parameters specified above could lead to a delay in surgery that...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-1808-2009
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Tenex Health
Product
American Optisurgical, Horizon Phacoemulsification System, Model: HZN-2030, for ophthalmic surgery.
Product code
HQC Unit, Phacofragmentation
Reason for recall
Device operation interrupted: A software bug causes Phaco to stop when vacuum goes above 300 mmHg while in Multi Burst mode. Use of this product at the parameters specified above could lead to a delay in surgery that might present a health risk.
Root cause tag
software
FDA root cause
Software change control
Action
American Optisurgery sent an E-mail to customers on April 2, 2009 stating the reason for recall and informed them to immediately examine their inventory and quarantine product subject to recall. In addition, if they have further distributed this product, please identify customers and identify them at once of this product recall. Their notification to customers may be enhanced by including a copy of this recall notification letter. They were then told to refer to the enclosed Recall Questionnaire for instructions on what to do with the recalled product: The recall should be carried out to the user level and return the enclosed response form as soon as possible. If there are any questions, they are advised to contact David Salzberg at (800) 576-1266.
Quantity in commerce
12 units
Distribution
Worldwide distribution: USA (California), Brazil, Algeria, France, South Korea, India, and Peru.
Product codes and lots
Serial Numbers: 1114, 1116, 1118, 1120-1124, and 1126-1129.