Portex Continuous Epidural Tray, Reference 4037-20 An anesthesia conduction kit is a device used to...
Smiths Medical ASD, Inc. began a recall of Portex Continuous Epidural Tray, Reference 4037-20 An anesthesia conduction kit is a device used to administer to a patient conduction, regional, or local... on . The reason FDA records is "Portex Continuous Epidural Tray was distributed with a 20G closed end, three eyed, nylon catheter without a stylet instead of the 20 G open end Teflon catheter with stylet specified on the label". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.
Recall record
- Recall number
- Z-2050-2009
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Abbott Medical
- Product
- Portex Continuous Epidural Tray, Reference 4037-20 An anesthesia conduction kit is a device used to administer to a patient conduction, regional, or local anesthesia.
- Product code
- BSO Catheter, Conduction, Anesthetic
- Reason for recall
- Portex Continuous Epidural Tray was distributed with a 20G closed end, three eyed, nylon catheter without a stylet instead of the 20 G open end Teflon catheter with stylet specified on the label.
- Root cause tag
- labeling and UDI
- FDA root cause
- Mixed-up of materials/components
- Action
- Smiths Medical notified consignees by an Urgent Device Recall letter dated July 15, 2009, to locate and quarantine the affected product, stop distribution, and complete and fax back the attached Distributor Reply Letter to Smiths Medical. Credit or replacement will be arranged as requested for the return of any affected product. All distributors will be required to return any affected product in stock. All of the distributors will be required to send their customer list to Smiths Medical who will notify their customers of this recall.
- Quantity in commerce
- 190 kits (19 cases)
- Distribution
- Nationwide Distribution -- including states of New York, Texas, Ohio, Utah, Alabama, Arizona, Kentucky, & Wyoming.
- Product codes and lots
- Lot # 1579258