Z-2039-2009

Advanced Bionics HiRes 90K Cochlear Implant, Model: CI-1400-01 and CI-1400-02H, Implant Magnet Case. Cochlear...

Advanced Bionics LLC / initiated 2009-06-10 / Terminated / root cause: manufacturing process

Advanced Bionics LLC began a recall of Advanced Bionics HiRes 90K Cochlear Implant, Model: CI-1400-01 and CI-1400-02H, Implant Magnet Case. Cochlear implant intended to restore a level of auditory... on . The reason FDA records is "The recall was initiated after Advanced Bionics became aware of fourteen (14) devices manufactured between May 27, 2005 and June 30, 2005 that are confirmed to have or potentially have an issue related to the internal...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-2039-2009
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Advanced Bionics LLC
Product
Advanced Bionics HiRes 90K Cochlear Implant, Model: CI-1400-01 and CI-1400-02H, Implant Magnet Case. Cochlear implant intended to restore a level of auditory sensation to patients with profound sensorineural deafness via electrical stimulation of the auditory nerve.
Product code
MCM Implant, Cochlear
Reason for recall
The recall was initiated after Advanced Bionics became aware of fourteen (14) devices manufactured between May 27, 2005 and June 30, 2005 that are confirmed to have or potentially have an issue related to the internal magnet. These devices are were made with an internal magnet where the metal case surrounding the magnet was not properly welded. Patients with these devices may experience a decreas
Root cause tag
manufacturing process
FDA root cause
Process control
Action
Advanced Bionics, LLC issued an "Important Notification" dated June 10, 2009 via Fed Ex to Consignees in the US and Canada. Outside of the US, the notification letters will be distributed in such a manner that delivery confirmation can be obtained. There were three type of notification letters distributed: one for Patients, one for Surgeons and one for Clinicians. Letters described the affected device prompting users on what to do next. An acknowledgement form was enclosed and should be completed and sent via fax to 1-661-362-7621. For further information, contact Advanced Bionics at 1-877-829-0026.
Quantity in commerce
18
Distribution
Worldwide Distribution -- US, Canada, France and South Korea.
Product codes and lots
Serial Numbers: 225271, 301184, 225466, 225647, 225289, 225187, 300922, 225629, 225512, 301129, 300926, 225272, 225373, 300980, 225281, 225423, 225540 and 301027.