BrightView Gamma Camera Radius Drive Assembly, manufactured by Philips Medical Systems (Cleveland) Inc...
Phillips Nuclear Medicine began a recall of BrightView Gamma Camera Radius Drive Assembly, manufactured by Philips Medical Systems (Cleveland) Inc., Cleveland, OH, 3/8" tilt, model number 4535604 34811... on . The reason FDA records is "Unexpected movement-- if the drive assembly is severely compromised, the detector may slide to its hardware limit". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1892-2009
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Philips North America LLC
- Product
- BrightView Gamma Camera Radius Drive Assembly, manufactured by Philips Medical Systems (Cleveland) Inc., Cleveland, OH, 3/8" tilt, model number 4535604 34811, 3/4"tilt, model number 4535674 34801, and 3/8" fixed, model number 4535604 34781, all catalog number 882480 Medical device-- intended to produce images depicting anatomical distributions of single photon emitting radioisotopes within the human body for interpretation by medical personnel.
- Product code
- KPS System, Tomography, Computed, Emission
- Reason for recall
- Unexpected movement-- if the drive assembly is severely compromised, the detector may slide to its hardware limit.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- Firm initiated its recall notification on July 6, 2009 issuing a field safety notice to its consignees via FedEx and internationally through its personnel. Responses will be tracked for effectiveness. Questions may be directed to Philips Customer Care Center at 1.800.722.9377.
- Quantity in commerce
- 288 units
- Distribution
- Worldwide
- Product codes and lots
- All units: 3/8" tilt, model number 4535604 34811, 3/4"tilt, model number 4535674 34801, and 3/8" fixed, model number 4535604 34781, all catalog number 882480