Z-1892-2009

BrightView Gamma Camera Radius Drive Assembly, manufactured by Philips Medical Systems (Cleveland) Inc...

Philips North America LLC / initiated 2009-07-02 / Terminated / root cause: manufacturing process

Phillips Nuclear Medicine began a recall of BrightView Gamma Camera Radius Drive Assembly, manufactured by Philips Medical Systems (Cleveland) Inc., Cleveland, OH, 3/8" tilt, model number 4535604 34811... on . The reason FDA records is "Unexpected movement-- if the drive assembly is severely compromised, the detector may slide to its hardware limit". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1892-2009
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Philips North America LLC
Product
BrightView Gamma Camera Radius Drive Assembly, manufactured by Philips Medical Systems (Cleveland) Inc., Cleveland, OH, 3/8" tilt, model number 4535604 34811, 3/4"tilt, model number 4535674 34801, and 3/8" fixed, model number 4535604 34781, all catalog number 882480 Medical device-- intended to produce images depicting anatomical distributions of single photon emitting radioisotopes within the human body for interpretation by medical personnel.
Product code
KPS System, Tomography, Computed, Emission
Reason for recall
Unexpected movement-- if the drive assembly is severely compromised, the detector may slide to its hardware limit.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
Firm initiated its recall notification on July 6, 2009 issuing a field safety notice to its consignees via FedEx and internationally through its personnel. Responses will be tracked for effectiveness. Questions may be directed to Philips Customer Care Center at 1.800.722.9377.
Quantity in commerce
288 units
Distribution
Worldwide
Product codes and lots
All units: 3/8" tilt, model number 4535604 34811, 3/4"tilt, model number 4535674 34801, and 3/8" fixed, model number 4535604 34781, all catalog number 882480