Stryker Orthopaedics Cancellous Bone Screw; Dia 5.5 mm; Lnth: 25 mm; Sterile: Made in the USA; Stryker...
Stryker Howmedica Osteonics Corp. began a recall of Stryker Orthopaedics Cancellous Bone Screw; Dia 5.5 mm; Lnth: 25 mm; Sterile: Made in the USA; Stryker Osteonics Corp. 325 Corporate Drive Mahwah, NJ 07430... on . The reason FDA records is "Raw material supplied by a third party used to manufacture various medical devices does not comply with metallurgical requirements outlines in ASTM standard for titanium surgical implants". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-2127-2009
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Howmedica Osteonics Corp
- Product
- Stryker Orthopaedics Cancellous Bone Screw; Dia 5.5 mm; Lnth: 25 mm; Sterile: Made in the USA; Stryker Osteonics Corp. 325 Corporate Drive Mahwah, NJ 07430. The bone screws are optional fixtures that may be used during knee and hip replacement procedures.
- Product code
- HWC Screw, Fixation, Bone
- Reason for recall
- Raw material supplied by a third party used to manufacture various medical devices does not comply with metallurgical requirements outlines in ASTM standard for titanium surgical implants.
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- Stryker Orthopaedics issued "Urgent Product Recall" letters via Federal Express priority overnight beginning June 26, 2009 to Branches, Hospital Risk Management and Chief of Orthopaedics and Surgeons. All users are requested to examine their inventory, retrieve all affected product and return for reconciliation to Stryker Orthopaedics. Users are also requested to complete the enclosed Product Accountability Form and fax with necessary information to the firm at 1-201-831-6069. For further information, contact Stryker Orthopaedics at 1-201-831-5970.
- Quantity in commerce
- 51,113 total - all sizes.
- Distribution
- Worldwide Distribution
- Product codes and lots
- Catalog number: 2029-5525-1; 5.5 CANCELLOUS BONE SCREW 25mm N0PMKE 24-Jun-2013 5.5 CANCELLOUS BONE SCREW 25mm N0NMKE 24-Jun-2013 5.5 CANCELLOUS BONE SCREW 25mm N0MMKE 24-Jun-2013 5.5 CANCELLOUS BONE SCREW 25mm N0RMKE 25-Jun-2013 5.5 CANCELLOUS BONE SCREW 25mm H1NMLE 21-Jul-2013 5.5 CANCELLOUS BONE SCREW 25mm JLWMLE 23-Jul-2013 5.5 CANCELLOUS BONE SCREW 25mm EL5MME 24-Aug-2013 5.5 CANCELLOUS BONE SCREW 25mm EMVMME 24-Aug-2013 5.5 CANCELLOUS BONE SCREW 25mm KW8MME 24-Aug-2013 5.5 CANCELLOUS BONE SCREW 25mm EL6MME 25-Aug-2013 5.5 CANCELLOUS BONE SCREW 25mm MER1AN 11-Nov-2013 5.5 CANCELLOUS BONE SCREW 25mm MER1AM 12-Nov-2013 5.5 CANCELLOUS BONE SCREW 25mm MET3JH 10-Dec-2013 5.5 CANCELLOUS BONE SCREW 25mm MET3JE 11-Dec-2013 5.5 CANCELLOUS BONE SCREW 25mm MET6KM 14-Dec-2013 5.5 CANCELLOUS BONE SCREW 25mm MET6AR 14-Dec-2013 5.5 CANCELLOUS BONE SCREW 25mm MET6KL 14-Dec-2013 5.5 CANCELLOUS BONE SCREW 25mm MET6LL 14-Dec-2013
FIRST ARTICLE / WEEKLY